The new high resolution ultra small-angle neutron scattering instrument at the High Flux Reactor in Grenoble

被引:59
作者
Hainbuchner, M [1 ]
Villa, M
Kroupa, G
Bruckner, G
Baron, M
Amenitsch, H
Seidl, E
Rauch, H
机构
[1] Osterreich Univ, Atominst, A-1020 Vienna, Austria
[2] Inst Max Von Laue Paul Langevin, F-38042 Grenoble, France
[3] Inst Biophys & Xray Struct Res, A-8010 Graz, Austria
关键词
D O I
10.1107/S0021889899013394
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
In 1998, the combined Interferometer and Ultra Small Angle Neutron Scattering (USANS) instrument S18 at the 58 MW High Flux Reactor at the Institute Laue-Langevin in Grenoble (France) started operation. The instrument has been upgraded to allow more advanced neutron optics experiments for fundamental, nuclear and condensed matter physics. The new supermirror guide along with the two channel cut silicon perfect crystals, configured as an advanced high resolution Bonse-Hart camera, provides optimal intensity conditions. A large range of wavelengths is accessible by variations of the Bragg angle and by using different reflecting planes of a properly cut silicon monochromator block in combination with different channel-cut analyzer crystals. The fine adjustment analyzer system is achieved by an advanced piezo drive, which has an accuracy of 0.036 seconds of arc. The basic features of the USANS camera, which takes advantage of a new tail suppression method, and the results of various test measurements concerning intensities, long term stability and the accessible wavelength range will be presented. A high resolution position sensitive detector can either be used to analyze beam profiles or to perform neutron radiography experiments.
引用
收藏
页码:851 / 854
页数:4
相关论文
共 10 条
[1]   Optimization of a Bonse-Hart Ultra-Small-Angle Neutron Scattering facility by elimination of the rocking-curve wings [J].
Agamalian, M ;
Wignall, GD ;
Triolo, R .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1997, 30 (pt 3) :345-352
[2]   Surface-induced parasitic scattering in Bonse-Hart double-crystal diffractometers [J].
Agamalian, M ;
Christen, DK ;
Drews, AR ;
Glinka, CJ ;
Matsuoka, H ;
Wignall, GD .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1998, 31 :235-240
[3]  
AMENTISCH H, 1999, UNPUB KERNTECHNIK, V64
[4]  
BIERMANN T, UNPUB
[5]   TAILLESS X-RAY SINGLE-CRYSTAL REFLECTION CURVES OBTAINED BY MULTIPLE REFLECTION - (SI GE CRYSTALS - E/T) [J].
BONSE, U ;
HART, M .
APPLIED PHYSICS LETTERS, 1965, 7 (09) :238-&
[6]   Diffraction enhanced x-ray imaging [J].
Chapman, D ;
Thomlinson, W ;
Johnston, RE ;
Washburn, D ;
Pisano, E ;
Gmur, N ;
Zhong, Z ;
Menk, R ;
Arfelli, F ;
Sayers, D .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (11) :2015-2025
[7]   Intensified CCDs as position sensitive neutron detectors [J].
Dietze, M ;
Felber, J ;
Raum, K ;
Rausch, C .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 377 (2-3) :320-324
[8]  
Guinier A., 1939, Ann Phys, V12, P161, DOI DOI 10.1051/ANPHYS/193911120161
[9]  
KROUPA G, UNPUB
[10]  
Podurets K. M., 1989, Soviet Physics - Technical Physics, V34, P654