Application of co-eluting structural analog internal standards for expanded linear dynamic range in liquid chromatography/electrospray mass spectrometry

被引:32
作者
Shi, GE [1 ]
机构
[1] Bristol Myers Squibb Co, Discovery Drug Metab & Pharmacokinet, Expt Stn, Wilmington, DE 19880 USA
关键词
D O I
10.1002/rcm.897
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Non-linear standard calibration curves occur frequently in liquid chromatography/electrospray ionization mass spectrometry (LC/ESI-MS), necessitating the use of non-linear regression curve fitting or a reduction in assay dynamic range. Non-linearity can be minimized by the use of stable isotope internal standards, if the cause of curve bending is ionization saturation. Unfortunately, stable isotope internal standards can be difficult or expensive to synthesize. Structural analog internal standards, if sufficiently close to analytes both in chemical structures and HPLC retention times, may mimic the effect of stable isotope internal standards. Tentative experimental evidence supporting this concept is presented in this paper. Copyright (C) 2002 John Wiley Sons, Ltd.
引用
收藏
页码:202 / 206
页数:5
相关论文
共 7 条
[1]  
[Anonymous], GUIDANCE IND BIOANAL
[2]   Determination of SR 49059 in human plasma and urine by LC-APCI/MS/MS [J].
Burton, R ;
Mummert, M ;
Newton, J ;
Brouard, R ;
Wu, DL .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1997, 15 (12) :1913-1922
[3]   CALIBRATION AND VALIDATION OF LINEARITY IN CHROMATOGRAPHIC BIOPHARMACEUTICAL ANALYSIS [J].
KARNES, HT ;
MARCH, C .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1991, 9 (10-12) :911-918
[4]  
Kebarle P. A., 1993, ANAL CHEM, V65, P972
[5]   Determination of capsaicin, dihydrocapsaicin, and nonivamide in self-defense weapons by liquid chromatography-mass spectrometry and liquid chromatography-tandem mass spectrometry [J].
Reilly, CA ;
Crouch, DJ ;
Yost, GS ;
Fatah, AA .
JOURNAL OF CHROMATOGRAPHY A, 2001, 912 (02) :259-267
[6]   Novel direct detection method for quantitative determination of intracellular nucleoside triphosphates using weak anion exchange liquid chromatography/tandem mass spectrometry [J].
Shi, G ;
Wu, JT ;
Li, Y ;
Geleziunas, R ;
Gallagher, K ;
Emm, T ;
Olah, T ;
Unger, S .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2002, 16 (11) :1092-1099
[7]   DEPENDENCE OF ION INTENSITY IN ELECTROSPRAY MASS-SPECTROMETRY ON THE CONCENTRATION OF THE ANALYTES IN THE ELECTROSPRAYED SOLUTION [J].
TANG, L ;
KEBARLE, P .
ANALYTICAL CHEMISTRY, 1993, 65 (24) :3654-3668