Six of Eight Hemoglobin A1c Point-of-Care Instruments Do Not Meet the General Accepted Analytical Performance Criteria

被引:171
作者
Lenters-Westra, Erna [1 ,2 ]
Slingerland, Robbert J. [1 ,2 ]
机构
[1] Isala Klin, Dept Clin Chem, NL-8011 JW Zwolle, Netherlands
[2] European Reference Lab Glycohemoglobin, NL-8011 JW Zwolle, Netherlands
关键词
GLYCOHEMOGLOBIN STANDARDIZATION PROGRAM;
D O I
10.1373/clinchem.2009.130641
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
100118 [医学信息学]; 100208 [临床检验诊断学];
摘要
BACKGROUND: Hemoglobin A(1c) (Hb A(1c)) point-of-care (POC) instruments are widely used to provide rapid-turnaround results in diabetic care centers. We investigated the conformance of various Hb A(1c) POC instruments (In2it from Bio-Rad, DCA Vantage from Siemens, Afinion and Nycocard from Axis-Shield, Clover from Infopia, InnovaStar from DiaSys, AICNow from Bayer, and Quo-Test from Quotient Diagnostics) with generally accepted performance criteria for Hb A(1c). METHODs: The CLSI protocols EP-10, EP-5, and EP-9 were applied to investigate imprecision, accuracy, and bias. We assessed bias using 3 certified secondary reference measurement procedures and the mean of the 3 reference methods. Assay conformance with the National Glycohemoglobin Standardization Program (NGSP) certification criteria, as calculated from analyses with 2 different reagent lot numbers for each Hb A, method, was also evaluated. RESULTS: Because of disappointing EP-10 results, 2 of the 8 manufacturers decided not to continue the evaluation. The total CVs from EP-5 evaluations for the different instruments with a low and high Hb A, value were: In2it 4.9% and 3.3%, DCA Vantage 1.8% and 3.7%, Clover 4.0% and 3.5%, InnovaStar 3.2% and 3.9%, Nycocard 4.8% and 5.2%, and Afinion 2.4% and 1.8%. Only the Afinion and the DCA Vantage passed the NGSP criteria with 2 different reagent lot numbers. CONCLUSIONS: Only the Afinion and the DCA Vantage met the acceptance criteria of having a total CV <3% in the clinically relevant range. The EP-9 results and the calculations of the NGSP certification showed significant differences in analytical performance be-tween different reagent lot numbers for all Hb A(1c) POC instruments. (C) 2009 American Association for Clinical Chemistry
引用
收藏
页码:44 / 52
页数:9
相关论文
共 25 条
[1]
Immediate feedback of HbA1c levels improves glycemic control in type I and insulin-treated type 2 diabetic patients [J].
Cagliero, E ;
Levina, EV ;
Nathan, DM .
DIABETES CARE, 1999, 22 (11) :1785-1789
[2]
TESTS OF EQUALITY BETWEEN SETS OF COEFFICIENTS IN 2 LINEAR REGRESSIONS [J].
CHOW, GC .
ECONOMETRICA, 1960, 28 (03) :591-605
[3]
*CLSI, 1992, EP5A NCCLS
[4]
*CLSI, 1998, EP10A CLSI
[5]
*CLSI, 1992, EP9A CLSI
[6]
Ferenczi A, 2001, Endocr Pract, V7, P85
[7]
Desirable performance standards for HbA1c analysis precision, accuracy and standardisation [J].
Goodall, Ian ;
Colman, Peter G. ;
Schneider, Hans G. ;
McLean, Mark ;
Barker, George .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2007, 45 (08) :1083-1097
[8]
Haemoglobin A1c:: Evaluation of three point of care analysers for use in a paediatric diabetes clinic [J].
Greaves, RF ;
Northfield, JAT ;
Cameron, FJ .
ANNALS OF CLINICAL BIOCHEMISTRY, 2005, 42 :124-129
[9]
Hawkins R. C., 2003, SMJ Singapore Medical Journal, V44, P008
[10]
HOMES EW, 2008, AM J CLIN PATHOL, V129, P540