The HG5 laser interferometer mercury manometer of the IMGC

被引:8
作者
Alasia, F [1 ]
Birello, G [1 ]
Capelli, A [1 ]
Cignolo, G [1 ]
Sardi, M [1 ]
机构
[1] CNR, Ist Metrol G Colonetti, I-10135 Turin, Italy
关键词
D O I
10.1088/0026-1394/36/6/3
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Mercury manometers have been designed and constructed at the Istituto di Metrologia "G. Colonnetti" (IMGC) since 1986, their main characteristic being the adoption of cube-corner reflectors, carried by lightweight floats, to reflect the beams of a laser interferometer that measures the vertical separation between the two mercury menisci. The paper describes the most recent instrument, designated HG5, designed to measure up to 120 kPa in either absolute or relative mode. The major improvements introduced in HG5 over its prototype (HG-3) are: the adoption of larger-bore columns; the installation of a single-beam laser interferometer that allows the use of smaller and less intrusive reflecting floats; the improvement of temperature control and measurement; and the development of cat's-eye floats of a novel design allowing the direct reflection of laser beams on mercury menisci. The uncertainty budget and the equations giving the HG5 standard uncertainties (43 mPa to 314 mPa) under all conditions are included.
引用
收藏
页码:499 / 503
页数:5
相关论文
共 10 条
[1]   The MM1 laser interferometer low-range mercury manometer of the IMGC [J].
Alasia, F ;
Capelli, A ;
Cignolo, G ;
Goria, R ;
Sardi, M .
METROLOGIA, 1999, 36 (06) :505-509
[2]  
Alasia F, 1993, METROLOGIA, V30, P571
[3]  
[Anonymous], PHYS METHODS CHEM
[4]  
CIGNOLO G, 1998, 26 IMGC
[5]  
CIGNOLO G, 1998, 14 IMGC
[6]  
CIGNOLO G, 1998, 15 IMGC
[7]  
PAVESE F, 1992, INT CRYOGENIC MONOGR, P280
[8]  
SOMMER KD, 1993, METROLOGIA, V30, P665
[9]   Improved measurement accuracy and uncertainty budget of the BIPM primary standard manobarometer [J].
Stock, M ;
Pello, R .
METROLOGIA, 1999, 36 (02) :149-153
[10]  