Geometry and Degree of Apposition of the CoreValve ReValving System With Multislice Computed Tomography After Implantation in Patients With Aortic Stenosis

被引:169
作者
Schultz, Carl J. [1 ]
Weustink, Annick [2 ]
Piazza, Nicolo [1 ]
Otten, Amber [1 ]
Mollet, Nico [2 ]
Krestin, Gabriel [2 ]
van Geuns, Robert J. [1 ]
de Feyter, Pim [1 ,2 ]
Serruys, Patrick W. J. [1 ]
de Jaegere, Peter [1 ]
机构
[1] Erasmus MC, Dept Cardiol, NL-3000 CA Rotterdam, Netherlands
[2] Erasmus MC, Dept Radiol, NL-3000 CA Rotterdam, Netherlands
关键词
cardiac computed tomography; percutaneous valve replacement; aortic valve; TRANSCATHETER IMPLANTATION; DISEASE; STENT;
D O I
10.1016/j.jacc.2009.04.075
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives Using multislice computed tomography (MSCT), we sought to evaluate the geometry and apposition of the CoreValve ReValving System (CRS, Medtronic, Luxembourgh, Luxembourgh) in patients with aortic stenosis. Background There are no data on the durability of percutaneous aortic valve replacement. Geometric factors may affect durability. Methods Thirty patients had MSCT at a median 1.5 months (interquartile range [IQR] 0 to 7 months) after percutaneous aortic valve replacement. Axial dimensions and apposition of the CRS were evaluated at 4 levels: 1) the ventricular end; 2) the nadir; 3) central coaptation of the CRS leaflets; and 4) commissures. Orthogonal smallest and largest diameters and cross-sectional surface area were measured at each level. Results The CRS (26-mm: n = 14, 29-mm: n = 16) was implanted at 8.5 mm (IQR 5.2 to 11.0 mm) below the noncoronary sinus. None of the CRS frames reached nominal dimensions. The difference between measured and nominal cross-sectional surface area at the ventricular end was 1.6 cm(2) (IQR 0.9 to 2.6 cm(2)) and 0.5 cm(2) (IQR 0.2 to 0.7 cm(2)) at central coaptation. At the level of central coaptation the CRS was undersized relative to the native annulus by 24% (IQR 15% to 29%). The difference between the orthogonal smallest and largest diameters (degree of deformation) at the ventricular end was 4.4 mm (IQR 3.3 to 6.4 mm) and it decreased progressively toward the outflow. Incomplete apposition of the CRS frame was present in 62% of patients at the ventricular end and was ubiquitous at the central coaptation and higher. Conclusions Dual-source MSCT demonstrated incomplete and nonuniform expansion of the CRS frame, but the functionally important mid-segment was well expanded and almost symmetrical. Undersizing and incomplete apposition were seen in the majority of patients. (J Am Coll Cardiol 2009; 54: 911-8) c 2009 by the American College of Cardiology Foundation
引用
收藏
页码:911 / 918
页数:8
相关论文
共 13 条
[1]   Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis - First human case description [J].
Cribier, A ;
Eltchaninoff, H ;
Bash, A ;
Borenstein, N ;
Tron, C ;
Bauer, F ;
Derumeaux, G ;
Anselme, F ;
Laborde, F ;
Leon, MB .
CIRCULATION, 2002, 106 (24) :3006-3008
[2]   Long-term outcomes after valve replacement for low-gradient aortic stenosis - Impact of prosthesis-patient mismatch [J].
Kulik, Alexander ;
Burwash, Ian G. ;
Kapila, Varun ;
Mesana, Thierry G. ;
Ruel, Marc .
CIRCULATION, 2006, 114 :I553-I558
[3]   Drug-eluting stent and coronary thrombosis -: Biological mechanisms and clinical implications [J].
Luescher, Thomas F. ;
Steffel, Jan ;
Eberli, Franz R. ;
Joner, Michael ;
Nakazawa, Gaku ;
Tanner, Felix C. ;
Virmani, Renu .
CIRCULATION, 2007, 115 (08) :1051-1058
[4]   Valvular aortic stenosis - Disease severity and timing of intervention [J].
Otto, Catherine M. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2006, 47 (11) :2141-2151
[5]  
Piazza Nicolo, 2008, EuroIntervention, V4, P242, DOI 10.4244/EIJV4I2A43
[6]   Anatomy of the Aortic Valvar Complex and Its Implications for Transcatheter Implantation of the Aortic Valve [J].
Piazza, Nicolo ;
de Jaegere, Peter ;
Schultz, Carl ;
Becker, Anton E. ;
Serruys, Patrick W. ;
Anderson, Robert H. .
CIRCULATION-CARDIOVASCULAR INTERVENTIONS, 2008, 1 (01) :74-81
[7]   Early and Persistent Intraventricular Conduction Abnormalities and Requirements for Pacemaking After Percutaneous Replacement of the Aortic Valve [J].
Piazza, Nicolo ;
Onuma, Yoshinobu ;
Jesserun, Emile ;
Kint, Peter Paul ;
Maugenest, Anne-Marie ;
Anderson, Robert H. ;
de Jaegere, Peter P. Th ;
Serruys, Patrick W. .
JACC-CARDIOVASCULAR INTERVENTIONS, 2008, 1 (03) :310-316
[8]  
SCHULTZ C, 2009, EUROINTERVE IN PRESS
[9]   THE DESIGN OF THE NORMAL AORTIC-VALVE [J].
THUBRIKAR, M ;
PIEPGRASS, WC ;
SHANER, TW ;
NOLAN, SP .
AMERICAN JOURNAL OF PHYSIOLOGY, 1981, 241 (06) :H795-H801
[10]   Noninvasive Evaluation of the Aortic Root With Multislice Computed Tomography Implications for Transcatheter Aortic Valve Replacement [J].
Tops, Laurens F. ;
Wood, David A. ;
Delgado, Victoria ;
Schuijf, Joanne D. ;
Mayo, John R. ;
Pasupati, Sanjeevan ;
Lamers, Frouke P. L. ;
van der Wall, Ernst E. ;
Schalij, Martin J. ;
Webb, John G. ;
Bax, Jeroen J. .
JACC-CARDIOVASCULAR IMAGING, 2008, 1 (03) :321-330