Analytical performance of the Investigational Use Only Cervista™ HPV HR test as determined by a multi-center study

被引:53
作者
Day, Stephen P. [1 ]
Hudson, Angela [1 ]
Mast, Andrea [1 ]
Sander, Tamara [1 ]
Curtis, Michelle [1 ]
Olson, Sarah [1 ]
Chehak, LuAnne [1 ]
Quigley, Neil [2 ]
Ledford, Joellen [2 ]
Yen-Lieberman, Belinda [3 ]
Kohn, Debra [3 ]
Quigley, Denise I. [4 ]
Olson, Marilyn [1 ]
机构
[1] Hologic Inc, Med Affairs, Madison, WI 53719 USA
[2] Mol Pathol Lab Network & Geneu CRS, Maryville, TN USA
[3] Cleveland Clin Fdn, Cleveland, OH 44195 USA
[4] Med Univ S Carolina, Dept Pathol & Lab Med, Charleston, SC 29425 USA
关键词
Cervista; Invader; Papillomavirus; HPV; Analytical; Performance; ATYPICAL SQUAMOUS CELLS; UNDETERMINED SIGNIFICANCE; HUMAN PAPILLOMAVIRUSES; CONSENSUS GUIDELINES; MANAGEMENT; DNA; CYTOLOGY; WOMEN;
D O I
10.1016/S1386-6532(09)70010-1
中图分类号
Q93 [微生物学];
学科分类号
071005 [微生物学];
摘要
Background: Any HPV test designed to be utilized in cervical cancer screening programs should be highly validated both analytically and clinically. Objectives: The Investigational Use Only (IUO) Cervista(TM) HPV FIR test is designed to detect 14 high-risk HPV types (16, 18, 31, 33, 35, 39, 45, 51, 52, 56, 58, 59, 66, and 68). The analytical performance of the Cervista(TM) HPV HR test was characterized in a multi-center study. Results: Analytical sensitivity for the 14 high-risk HPV types that the test is designed to detect ranged from 1,250 copies to 7,500 copies per reaction depending on HPV type. Accuracy compared to PCR with bi-directional sequencing was 91.4% [95% CI: 86.5-95.0%]. The reproducibility, when tested at three different testing centers, resulted in an overall inter-run reproducibility (between day/within site) agreement of 98.8% [1-sided 95% Confidence Lower Limit = 96.9%] and an overall inter-site reproducibility (between site) agreement of 98.7% [1-sided 95% Confidence Lower Limit = 97.9%]. The Cervista(TM) HPV HR test showed no cross-reactivity with DNA from seven non-oncogenic HPV types or 17 different infectious agents at up to 10(7) copies per reaction. Conclusions: The analytical performance of the Cervista(TM) HPV HR test demonstrates sufficient anatytical performance for use in cervical cancer screening. As with any clinical. laboratory test, anatytical. characteristics must be evaluated in tight of the clinical performance of this assay. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:S63 / S72
页数:10
相关论文
共 21 条
[1]
[Anonymous], 2005, IARC HDB CANC, V10
[2]
Carcinogenicity of human papillomaviruses [J].
Cogliano, V ;
Baan, R ;
Straif, K ;
Grosse, Y ;
Secretan, B ;
El Ghissassi, F .
LANCET ONCOLOGY, 2005, 6 (04) :204-204
[3]
Davey DD, 2004, ARCH PATHOL LAB MED, V128, P1224
[4]
DAY SP, 2005, ENCY DIAGNOSTIC GENO, P682
[5]
de Arruda Monika, 2002, Expert Rev Mol Diagn, V2, P487
[6]
Gravitt PE, 2000, J CLIN MICROBIOL, V38, P357
[7]
Clinical utility of HPV genotyping [J].
Meijer, Chris J. ;
Snijders, Peter J. ;
Castle, Philip E. .
GYNECOLOGIC ONCOLOGY, 2006, 103 (01) :12-17
[8]
NOLTE FS, 2006, CLSI GUIDELINE
[9]
Poljak M, 2002, J CLIN VIROL, V25, pS89
[10]
American Cancer Society guideline for the early detection of cervical neoplasia and cancer [J].
Saslow, D ;
Runowicz, CD ;
Solomon, D ;
Moscicki, AB ;
Smith, RA ;
Eyre, HJ ;
Cohen, C .
CA-A CANCER JOURNAL FOR CLINICIANS, 2002, 52 (06) :342-362