Clinical predictors of transvenous biphasic defibrillation thresholds

被引:34
作者
Gold, MR [1 ]
Khalighi, K [1 ]
Kavesh, NG [1 ]
Daly, B [1 ]
Peters, RW [1 ]
Shorofsky, SR [1 ]
机构
[1] UNIV MARYLAND MED SYST,DEPT MED,DIV CARDIOL,BALTIMORE,MD
关键词
D O I
10.1016/S0002-9149(97)00210-5
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Transvenous lead systems have become routine for defibrillator placement. However, previous studies of clinical predictors of an adequate nonthoracotomy defibrillation threshold (DFT) evaluated monophasic waveforms or more complex lead systems, including subcutaneous patches. Accordingly, this study is a prospective evaluation of the predictors of an adequate biphasic DTT in 114 consecutive patients undergoing cardioverter-defibrillator implantation with a single transvenous lead. For each subject, 38 parameters were assessed, including standard demographic, electrocardiographic, echocardiographic, and radiographic measurements. An adequate DFT (less than or equal to 20 J) was achieved in 92% of patients. Multivariable analysis revealed 2 independent factors predictive of a high threshold: echocardiographic measurements of left ventricular dilation (odds ratio = 0.16, 95% confidence interval 0.05 to 0.53, p = 0.003) and body size (odds ratio = 0.36, 95% confidence interval 0.17 to 0.73; p = 0.005). No patient with a normal left ventricular end-diastolic dimension had a high DFT, whereas 14% (9 of 66) of those with left ventricular dilation had elevated thresholds. When the DFT cutoff was lowered to 15 J, as is necessary with some downsized pulse generators, an adequate threshold was observed in 84% of patients and the same 2 independent predictors of high thresholds were found. These results indicate that an adequate transvenous DFT can be predicted from simple clinical parameters. (C) 1997 by Excerpta Medica, Inc.
引用
收藏
页码:1623 / 1627
页数:5
相关论文
共 30 条
[1]   A SIMPLIFIED, SINGLE-LEAD UNIPOLAR TRANSVENOUS CARDIOVERSION-DEFIBRILLATION SYSTEM [J].
BARDY, GH ;
JOHNSON, G ;
POOLE, JE ;
DOLACK, GL ;
KUDENCHUK, PJ ;
KELSO, D ;
MITCHELL, R ;
MEHRA, R ;
HOFER, B .
CIRCULATION, 1993, 88 (02) :543-547
[2]   A PROSPECTIVE RANDOMIZED COMPARISON IN HUMANS OF BIPHASIC WAVE-FORM 60-MU-F AND 120-MU-F CAPACITANCE PULSES USING A UNIPOLAR DEFIBRILLATION SYSTEM [J].
BARDY, GH ;
POOLE, JE ;
KUDENCHUK, PJ ;
DOLACK, GL ;
MEHRA, R ;
DEGROOT, P ;
RAITT, MH ;
JONES, GK ;
JOHNSON, G .
CIRCULATION, 1995, 91 (01) :91-95
[3]   RESULTS AND REALISTIC EXPECTATIONS WITH TRANSVENOUS LEAD SYSTEMS [J].
BLOCK, M ;
HAMMEL, D ;
ISBRUCH, F ;
BORGGREFE, M ;
WIETHOLT, D ;
HACHENBERG, T ;
SCHELD, HH ;
BREITHARDT, G .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (04) :665-670
[4]   A PROSPECTIVE RANDOMIZED CROSS-OVER COMPARISON OF MONOPHASIC AND BIPHASIC DEFIBRILLATION USING NONTHORACOTOMY LEAD CONFIGURATIONS IN HUMANS [J].
BLOCK, M ;
HAMMEL, D ;
BOCKER, D ;
BORGGREFE, M ;
BUDDE, T ;
ISBRUCH, F ;
WIETHOLT, D ;
SCHELD, HH ;
BREITHARDT, G .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1994, 5 (07) :581-590
[5]   INTERNAL DEFIBRILLATION WITH SMALLER CAPACITORS - A PROSPECTIVE RANDOMIZED CROSS-OVER COMPARISON OF DEFIBRILLATION EFFICACY OBTAINED WITH 90-MU-F AND 125-MU-F CAPACITORS IN HUMANS [J].
BLOCK, M ;
HAMMEL, D ;
BOCKER, D ;
BORGGREFE, M ;
SEIFERT, T ;
FASTENRATH, C ;
SCHELD, HH ;
BREITHARDT, G .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1995, 6 (05) :333-342
[6]  
BROOKS R, 1993, J AM COLL CARDIOL, V22, P1833
[7]   ECHOCARDIOGRAPHIC DETERMINATION OF LEFT-VENTRICULAR MASS IN MAN - ANATOMIC VALIDATION OF METHOD [J].
DEVEREUX, RB ;
REICHEK, N .
CIRCULATION, 1977, 55 (04) :613-618
[8]   Transvenous defibrillation lead systems [J].
Gold, MR ;
Shorofsky, SR .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1996, 7 (06) :570-580
[9]   Effects of an active pectoral-pulse generator shell on defibrillation efficacy with a transvenous lead system [J].
Gold, MR ;
Foster, AH ;
Shorofsky, SR .
AMERICAN JOURNAL OF CARDIOLOGY, 1996, 78 (05) :540-543
[10]   Clinical predictors of transvenous defibrillation energy requirements [J].
Khalighi, K ;
Daly, B ;
Leino, EV ;
Shorofsky, SR ;
Kavesh, NG ;
Peters, RW ;
Gold, MR .
AMERICAN JOURNAL OF CARDIOLOGY, 1997, 79 (02) :150-153