Monitoring gastric mucosal carbon dioxide pressure using gas tonometry: In vitro and in vivo validation studies

被引:82
作者
Creteur, J [1 ]
DeBacker, D [1 ]
Vincent, JL [1 ]
机构
[1] FREE UNIV BRUSSELS,ERASME UNIV HOSP,DEPT INTENS CARE,B-1070 BRUSSELS,BELGIUM
关键词
gastric intramucosal carbon dioxide pressure; monitoring; splanchnic; gastric tonometry; gas tonometry; saline tonometry oxygen delivery; splanchnic ischemia;
D O I
10.1097/00000542-199709000-00008
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: Saline gastric tonometry of carbon dioxide has been proposed as a means to assess the adequacy of splanchnic perfusion. However, this technique has several disadvantages, including the long time interval needed for gases to reach equilibrium in saline milieu. Thus the authors evaluated a system that uses a gas-filled instead of a saline-filled gastric balloon. Methods: In vitro, we simultaneously placed two tonometry catheters in an equilibration water bath maintained at a predetermined and constant pressure of carbon dioxide (P-co2). The first catheter's balloon was filled with air and the second with saline. The performance of gas tonometry was tested by comparing the P-co2 measurements of the bath obtained via gas tonometry (Pg(co2)) to the P-co2 measurements of direct bath samples (Pbath(co2)). These results were also compared with the P-co2 measurements obtained simultaneously by saline tonometry (Ps(co2)). The response time of gas versus saline tonometry was also studied. In vivo, the performance of gas tonometry was tested comparing the measurements of gastric intramucosal P-co2 obtained by gas tonometry (Pg(co2)) at different equilibration times with those obtained by saline tonometry (Ps(co2)) using an equilibration time of 30 min. Two nasogastric tonometry catheters were placed simultaneously in seven stable patients in the intensive care unit. The first balloon was filled with air and the second with saline. Results: In vitro, there was a close correlation between Pg(co2) and Pbath(co2), for each level of Pbath(co2), and for each different gas equilibration time. For an equilibration time of 10 min at a Pbath(co2) level of approximately 40 mmHg, the bias of the gas device defined as the mean of the differences between Pbath(co2) and Pg(co2), and its precision defined as the standard deviation of the bias, were -0.3 mmHg and 0.7 mmHg, respectively. Using the same definitions, the bias and precision of saline tonometry were 11.2 mmHg and 1.4 mmHg, respectively. If the equilibration time-dependent correction factor provided by the catheter manufacturer for saline tonometry was applied, the bias and precision were -6.9 mmHg and 2.9 mmHg, respectively. In vivo, using an equilibration time of 10 min for gas and 30 min for saline tonometry, there was a close correlation between the two techniques (r(2) = 0.986). A Bland and Altman analysis revealed a bias (+/- 2 SD) of 0.1 +/- 6.8 mmHg. The correlation between the two methods was not improved if we prolonged the equilibration time of the gas tonometer. Conclusions: Gas tonometry is comparable to saline tonometry for measuring gastric intramucosal P-co2. Because gas tonometry is easier to automate, it may offer advantages over saline tonometry.
引用
收藏
页码:504 / 510
页数:7
相关论文
共 29 条
  • [2] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [3] BODA D, 1959, LANCET, V1, P181
  • [4] COMPARISON OF CLINICAL INFORMATION GAINED FROM ROUTINE BLOOD-GAS ANALYSIS AND FROM GASTRIC TONOMETRY FOR INTRAMURAL PH
    BOYD, O
    MACKAY, CJ
    LAMB, G
    BLAND, JM
    GROUNDS, RM
    BENNETT, ED
    [J]. LANCET, 1993, 341 (8838) : 142 - 146
  • [5] GASTRIC TONOMETRY SUPPLEMENTS INFORMATION PROVIDED BY SYSTEMIC INDICATORS OF OXYGEN-TRANSPORT
    CHANG, MC
    CHEATHAM, ML
    NELSON, LD
    RUTHERFORD, EJ
    MORRIS, JA
    [J]. JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1994, 37 (03) : 488 - 494
  • [6] GASTRIC-MUCOSAL PH AS A PROGNOSTIC INDEX OF MORTALITY IN CRITICALLY ILL PATIENTS
    DOGLIO, GR
    PUSAJO, JF
    EGURROLA, MA
    BONFIGLI, GC
    PARRA, C
    VETERE, L
    HERNANDEZ, MS
    FERNANDEZ, S
    PALIZAS, F
    GUTIERREZ, G
    [J]. CRITICAL CARE MEDICINE, 1991, 19 (08) : 1037 - 1040
  • [7] BACK-DIFFUSION OF CO2 AND ITS INFLUENCE ON THE INTRAMURAL PH IN GASTRIC-MUCOSA
    FIDDIANGREEN, RG
    PITTENGER, G
    WHITEHOUSE, WM
    [J]. JOURNAL OF SURGICAL RESEARCH, 1982, 33 (01) : 39 - 48
  • [9] COMBINED MEASUREMENTS OF BLOOD LACTATE CONCENTRATIONS AND GASTRIC INTRAMUCOSAL PH IN PATIENTS WITH SEVERE SEPSIS
    FRIEDMAN, G
    BERLOT, G
    KAHN, RJ
    VINCENT, JL
    [J]. CRITICAL CARE MEDICINE, 1995, 23 (07) : 1184 - 1193
  • [10] A TRIAL OF GOAL-ORIENTED HEMODYNAMIC THERAPY IN CRITICALLY ILL PATIENTS
    GATTINONI, L
    BRAZZI, L
    PELOSI, P
    LATINI, R
    TOGNONI, G
    PESENTI, A
    FUMAGALLI, R
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1995, 333 (16) : 1025 - 1032