Evaluation of hemolysis in the VentrAssist implantable rotary blood pump

被引:42
作者
James, NL [1 ]
Wilkinson, CM [1 ]
Lingard, NL [1 ]
van der Meer, AL [1 ]
Woodard, JC [1 ]
机构
[1] Ventracor Ltd, VentrAssist Div, Sydney, NSW, Australia
关键词
phosphate buffered saline; hemaccel; hemolysis; normalized index of hemolysis; surface finish;
D O I
10.1046/j.1525-1594.2003.07188.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The VentrAssist implantable rotary blood pump (IRBP) is an implantable centrifugal blood pump with a hydrodynamically suspended impeller; optimal efficiency requires small running clearances (70-300 mum). The effect of running clearance and polish on hemolysis was evaluated in vitro. Three different human blood suspensions were compared: phosphate buffered saline (PBS), plasma volume expander (Hemaccel), and whole blood. The test conditions were: blood hematocrit 30%, flow rate 5 L/min, pressure across pump 100 mm Hg, 6 h flow period, and 37degreesC. Normalized Index of Hemolysis (NIH) for the Biomedicus BP-80, used as a control, was: 0.0040 +/- 0.0023 (n = 9; x +/- SD) and 0.00014 +/- 0.00009 (n = 5) for pooled blood suspensions in PBS and Hemaccel respectively, and 0.00053 +/- 0.0002 (n = 3) in whole blood. Hemolysis was reduced by improved surface finish and unaffected by running clearance. NIH for the VentrAssist IRBP with 0.2 mum Ra surface finish was 0.000167 +/- 0.00007 (n = 4) g/100 L in whole human blood, demonstrating minimal hemolysis.
引用
收藏
页码:108 / 113
页数:6
相关论文
共 10 条
[1]  
[Anonymous], F184197 ASTM
[2]  
[Anonymous], F183097 ASTM
[3]   The suspension stability of the blood [J].
Fahraeus, R .
PHYSIOLOGICAL REVIEWS, 1929, 9 (02) :241-274
[5]   COMPARATIVE HEMOLYSIS TESTS OF ROTARY BLOOD PUMPS [J].
ITKIN, GP ;
MATVEEV, YG ;
ROMANOV, O .
ARTIFICIAL ORGANS, 1995, 19 (07) :616-619
[6]   Species differences in erythrocyte mechanical fragility - Comparison of human, bovine, and ovine cells [J].
Jikuya, T ;
Tsutsui, T ;
Shigeta, O ;
Sankai, Y ;
Mitsui, T .
ASAIO JOURNAL, 1998, 44 (05) :M452-M455
[7]   DOES HEMATOCRIT AFFECT IN-VITRO HEMOLYSIS TEST-RESULTS - PRELIMINARY-STUDY WITH BAYLOR/NASA PROTOTYPE AXIAL-FLOW PUMP [J].
MIZUGUCHI, K ;
DAMM, GA ;
ABER, GS ;
BOZEMAN, RJ ;
BACAK, JW ;
SVEJKOVSKY, PA ;
ORIME, Y ;
OHARA, Y ;
NAITO, K ;
TASAI, K ;
MAKINOUCHI, K ;
TAKATANI, S ;
NOSE, Y ;
NOON, GP ;
DEBAKEY, ME .
ARTIFICIAL ORGANS, 1994, 18 (09) :650-656
[8]  
Takami Y, 1996, ARTIF ORGANS, V20, P1155
[9]   VentrAssist hydrodynamically suspended, open, centrifugal blood pump [J].
Watterson, PA ;
Woodard, JC ;
Ramsden, VS ;
Reizes, JA .
ARTIFICIAL ORGANS, 2000, 24 (06) :475-477
[10]   A MATHEMATICAL-MODEL FOR SHEAR-INDUCED HEMOLYSIS [J].
YELESWARAPU, KK ;
ANTAKI, JF ;
KAMENEVA, MV ;
RAJAGOPAL, KR .
ARTIFICIAL ORGANS, 1995, 19 (07) :576-582