Gunther Tulip retrievable inferior vena caval filters: Indications, efficacy, retrieval, and complications

被引:70
作者
Looby, S. [1 ]
Given, M. F. [1 ]
Geoghegan, T. [1 ]
McErlean, A. [1 ]
Lee, M. J. [1 ]
机构
[1] Beaumont Hosp, Dept Radiol, Dublin 9, Ireland
关键词
filter design; filter retrieval; Gunther Tulip filter; pulmonary embolism; retrieval interval; thromboembolic disease;
D O I
10.1007/s00270-006-0093-9
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Purpose: We evaluated the Gunther Tulip (GT) retrievable inferior vena cava (IVC) filter with regard to indications, filtration efficacy, complications, retrieval window, and use of anticoagulation. Methods: A retrospective study was performed of 147 patients (64 men, 83 women; mean age 58.8 years) who underwent retrievable GT filter insertion between 2001 and 2005. The indications for placement included a diagnosis of pulmonary embolism or deep venous thrombosis with a contraindication to anticoagulation (n = 68), pulmonary embolism or deep venous thrombosis while on anticoagulation (n = 49), prophylactic filter placement for high-risk surgical patients with a past history of pulmonary embolism or deep venous thrombosis (n = 20), and a high risk of pulmonary embolism or deep venous thrombosis (n = 10). Forty-nine of the 147 patients did not receive anticoagulation (33.7%) while 96 of 147 patients did, 82 of these receiving warfarin (56.5%), 11 receiving low-molecular weight heparins (7.58%), and 3 receiving antiplatelet agents alone (2.06%). Results: Filter placement was successful in 147 patients (100%). Two patients had two filters inserted. Of the 147 patients, filter deployment was on a permanent basis in 102 and with an intention to retrieve in 45 patients. There were 36 (80%) successful retrievals and 9 (20%) failed retrievals. The mean time to retrieval was 33.6 days. The reasons for failed retrieval included filter struts tightly adherent to the IVC wall (5/9), extreme filter tilt (2/9), and extensive filter thrombus (2/9). Complications included pneumothorax (n = 4), failure of filter expansion (n = 1), and breakthrough pulmonary embolism (n = 1). No IVC thrombotic episodes were recorded. Discussion: The Gunther Tulip retrievable filter can be used as a permanent or a retrievable filter. It is safe and efficacious. GT filters can be safely retrieved at a mean time interval of 33.6 days. The newly developed Celect filter may extend the retrieval interval.
引用
收藏
页码:59 / 65
页数:7
相关论文
共 28 条
[1]  
[Anonymous], 1998, Thrombosis and Emboli
[2]   Initial experience in humans with a new retrievable inferior vena cava filter [J].
Asch, MR .
RADIOLOGY, 2002, 225 (03) :835-844
[3]   Inferior vena caval filters: Review of a 26-year single-center clinical experience [J].
Athanasoulis, CA ;
Kaufman, JA ;
Halpern, EF ;
Waltman, AC ;
Geller, SC ;
Fan, CM .
RADIOLOGY, 2000, 216 (01) :54-66
[4]   Long-term follow-up of Vena Tech-LGM filter:: Predictors and frequency of caval occlusion [J].
Crochet, DP ;
Brunel, P ;
Trogrlic, S ;
Grossetête, R ;
Auget, JL ;
Dary, C .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1999, 10 (02) :137-142
[5]   A clinical trial of vena caval filters in the prevention of pulmonary embolism in patients with proximal deep-vein thrombosis [J].
Decousus, H ;
Leizorovicz, A ;
Parent, F ;
Page, Y ;
Tardy, B ;
Girard, P ;
Laporte, S ;
Faivre, R ;
Charbonnier, B ;
Barral, FG ;
Huet, Y ;
Simonneau, G .
NEW ENGLAND JOURNAL OF MEDICINE, 1998, 338 (07) :409-415
[6]   Medical literature and vena cava filters - So far so weak [J].
Girard, P ;
Stern, JB ;
Parent, F .
CHEST, 2002, 122 (03) :963-967
[7]   Experience with the recovery filter as a retrievable inferior vena cava filter [J].
Grande, WJ ;
Trerotola, SO ;
Reilly, PM ;
Clark, TWI ;
Soulen, MC ;
Patel, A ;
Shlansky-Goldberg, RD ;
Tuite, CM ;
Solomon, JA ;
Mondschein, JI ;
Fitzpatrick, MK ;
Stavropoulos, SW .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2005, 16 (09) :1189-1193
[8]   The percutaneous Greenfield filter: Outcomes and practice patterns - Discussion [J].
Rutherford, R ;
Greenfield, LJ .
JOURNAL OF VASCULAR SURGERY, 2000, 32 (05) :893-893
[9]   New optional IVC filter for percutaneous retrieval -: In vitro evaluation of embolus capturing efficiency [J].
Günther, RW ;
Neuerburg, J ;
Mossdorf, A ;
Pfeffer, J ;
Hoj, AR ;
Molgaard-Nielsen, A ;
Bücker, A ;
Schmitz-Rode, T .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2005, 177 (05) :632-636
[10]   The role of vena caval filters in the management of venous thromboembolism [J].
Hann, CL ;
Streiff, MB .
BLOOD REVIEWS, 2005, 19 (04) :179-202