Integrated portable polymerase chain reaction-capillary electrophoresis microsystem for rapid forensic short tandem repeat typing

被引:141
作者
Liu, Peng
Seo, Tae Seok
Beyor, Nathaniel
Shin, Kyoung-Jin
Scherer, James R.
Mathies, Richard A. [1 ]
机构
[1] UCSF, UC Berkeley Joint Grad Grp Bioengn, Berkeley, CA 94720 USA
[2] Univ Calif Berkeley, Dept Chem, Berkeley, CA 94720 USA
关键词
D O I
10.1021/ac061961k
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A portable forensic genetic analysis system consisting of a microfluidic device for amplification and separation of short tandem repeat (STR) fragments as well as an instrument for chip operation and four-color fluorescence detection has been developed. The microdevice performs polymerase chain reaction (PCR) in a 160-nL chamber and capillary electrophoresis (CE) in a 7-cm-long separation channel. The instrumental design integrates PCR thermal cycling, electrophoretic separation, pneumatic valve fluidic control, and four-color laser excited fluorescence detection. A quadruplex Y-chromosome STR typing system consisting of amelogenin and three Y STR loci (DYS390, DYS393, DYS439) was developed and used for validation studies. The multiplex amplification of these 4 loci with 35 PCR cycles followed by CE separation and 4-color fluorescence detection was completed in 1.5 h. All the amplicons can be detected with a limit of detection of 20 copies of male standard DNA in the reactor. Real-world forensic analyses of oral swab and human bone extracts from case evidence were also successfully performed. Mixture analysis demonstrated that a balanced profile can be obtained even at a male-to-female template ratio of 1:10. The successful development and operation of this portable PCR-CE system establishes the feasibility of rapid point-of-analysis DNA typing of forensic casework, of mass disaster samples or of individuals at a security checkpoint.
引用
收藏
页码:1881 / 1889
页数:9
相关论文
共 42 条
[1]   POSTMORTEM STABILITY OF DNA [J].
BAR, W ;
KRATZER, A ;
MACHLER, M ;
SCHMID, W .
FORENSIC SCIENCE INTERNATIONAL, 1988, 39 (01) :59-70
[2]  
Belgrader P, 1998, J FORENSIC SCI, V43, P315
[3]   Microchip-based cell lysis and DNA extraction from sperm cells for application to forensic analysis [J].
Bienvenue, JM ;
Duncalf, N ;
Marchiarullo, D ;
Ferrance, JP ;
Landers, JP .
JOURNAL OF FORENSIC SCIENCES, 2006, 51 (02) :266-273
[4]   Microfabricated bioprocessor for integrated nanoliter-scale Sanger DNA sequencing [J].
Blazej, Robert G. ;
Kumaresan, Palani ;
Mathies, Richard A. .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2006, 103 (19) :7240-7245
[5]   Microchip-based purification of DNA from biological samples [J].
Breadmore, MC ;
Wolfe, KA ;
Arcibal, IG ;
Leung, WK ;
Dickson, D ;
Giordano, BC ;
Power, ME ;
Ferrance, JP ;
Feldman, SH ;
Norris, PM ;
Landers, JP .
ANALYTICAL CHEMISTRY, 2003, 75 (08) :1880-1886
[6]  
Burger MF, 2005, BIOTECHNIQUES, V39, P257
[7]  
Butler J. M., 2005, FORENSIC STR TYPING
[8]   Genetics and genomics of core short tandem repeat loci used in human identity testing [J].
Butler, JM .
JOURNAL OF FORENSIC SCIENCES, 2006, 51 (02) :253-265
[9]  
Chakraborty R, 1999, ELECTROPHORESIS, V20, P1682
[10]   Control and detection of chemical reactions in microfluidic systems [J].
deMello, Andrew J. .
NATURE, 2006, 442 (7101) :394-402