Glow discharge plasma treatment of polyethylene tubing with tetraglyme results in ultralow fibrinogen adsorption and greatly reduced platelet adhesion

被引:41
作者
Cao, Lan
Sukavaneshvar, Sivaprasad
Ratner, Buddy D.
Horbett, Thomas A. [1 ]
机构
[1] Univ Washington, Dept Chem Engn, Seattle, WA 98195 USA
[2] Utah Artificial Heart Inst, Salt Lake City, UT 84103 USA
[3] Univ Washington, Dept Bioengn, Seattle, WA 98195 USA
关键词
hemocompatibility; glow discharge plasma deposition; protein adsorption; platelet adhesion; tetraglyme;
D O I
10.1002/jbm.a.30908
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Previous studies from our lab have shown that fibrinogen adsorption (Gamma(Fg)) must be reduced below 10 ng/cm(2) to significantly reduce platelet adhesion, and that radio frequency glow discharge (RFGD) treatment of polymeric films in the presence of tetraethylene glycol dimethyl ether (tetraglyme) can reduce Gamma(Fg) to the desired ultralow value. In this report, the effects of RFGD coatings of tetraglyme on the lumenal surface of PE tubing on Gamma(Fg) and on blood interactions both in vitro and ex vivo are described. Gamma(Fg) on the tetraglyme-coated PE tubing was reduced to the desired ultralow level (< 10 ng/cm(2)), and we also observed a significant decrease in adsorption of von Willebrand's factor. In vitro platelet adhesion from washed platelet suspensions, platelet rich plasma, or whole blood to tetraglyme-coated PE tubing was decreased compared to PE, polyurethane, or silicone rubber tubes. In addition, thrombin generation by platelets adherent to tetraglyme-coated PE was also much less than by platelets adherent to PE. When inserted in an ex vivo carotid artery-carotid artery shunt in sheep, the RFGD tetraglyme-coated PE exhibited a very low number of adherent platelets compared to heparin-coated, chromic acid-etched, or plain PE. The RFGD tetraglyme-coated PE tubes exhibited high protein and platelet resistance in vitro, and high platelet resistance ex vivo. The improved hemocompatibility is attributed to the unique chemical structure of RFGD tetraglyme that makes it highly protein resistant. (c) 2006 Wiley Periodicals, Inc.
引用
收藏
页码:788 / 803
页数:16
相关论文
共 69 条
[1]   Surface modification of PET films using pulsed AC plasma polymerisation aimed at preventing protein adsorption [J].
Ademovic, Z ;
Wei, J ;
Winther-Jensen, B ;
Hou, XL ;
Kingshott, P .
PLASMA PROCESSES AND POLYMERS, 2005, 2 (01) :53-63
[2]  
[Anonymous], BIOMATERIALS SCI INT
[3]  
Beyer D, 1997, J BIOMED MATER RES, V36, P181, DOI 10.1002/(SICI)1097-4636(199708)36:2<181::AID-JBM6>3.0.CO
[4]  
2-G
[5]   SMALL DIAMETER BLOOD-VESSEL PROSTHESES FROM BLENDS OF POLYETHYLENE OXIDE AND POLYPROPYLENE OXIDE [J].
BOTS, JGF ;
VANDERDOES, L ;
BANTJES, A .
BIOMATERIALS, 1986, 7 (05) :393-399
[6]  
CAO L, IN PRESS J BIOMED MA
[7]   PEO ENHANCEMENT OF PLATELET DEPOSITION, FIBRINOGEN DEPOSITION, AND COMPLEMENT C3 ACTIVATION [J].
CHAIKOF, EL ;
MERRILL, EW ;
CALLOW, AD ;
CONNOLLY, RJ ;
VERDON, SL ;
RAMBERG, K .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1992, 26 (09) :1163-1168
[8]   Cytokine release and neutrophil activation are not prevented by heparin-coated circuits and aprotinin administration [J].
Defraigne, JO ;
Pincemail, J ;
Larbuisson, R ;
Blaffart, F ;
Limet, R .
ANNALS OF THORACIC SURGERY, 2000, 69 (04) :1084-1091
[9]   SOLUTION TECHNIQUE TO INCORPORATE POLYETHYLENE OXIDE AND OTHER WATER-SOLUBLE POLYMERS INTO SURFACES OF POLYMERIC BIOMATERIALS [J].
DESAI, NP ;
HUBBELL, JA .
BIOMATERIALS, 1991, 12 (02) :144-153
[10]  
Dewanjee MK, 1999, J HEART VALVE DIS, V8, P324