Determination of low analyte concentrations by near-infrared spectroscopy: Effect of spectral pretreatments and estimation of multivariate detection limits

被引:58
作者
Blanco, Marcel [1 ]
Castillo, Miguel
Peinado, Antonio
Beneyto, Rafael
机构
[1] Univ Autonoma Barcelona, Fac Sci, Dept Chem, E-08193 Barcelona, Spain
[2] ICI Spain SA, Dev & Qual Assurance Dept, Barcelona 08495, Spain
关键词
near infrared spectroscopy; detection limit; orthogonal signal correction; net analyte signal; low concentration analytes;
D O I
10.1016/j.aca.2006.08.018
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Near infrared spectroscopy (NIRS) was used in combination with partial least squares (PLS) calibration to determine low concentrated anallytes. The effect of the orthogonal signal correction (OSC) and net analyte signal (NAS) pretreatments on the models obtained at concentrations of analyte near its detection limit was studied. Both pretreatments were found to accurately resolve the analyte signal and allow the construction of PLS models from a reduced number of factors; however, they provided no substantial advantage in terms of %RSE for the prediction samples. Multiple methodologies for the estimation of detection limits could be found in the bibliography. Nevertheless, detection limits were determined by a multivariate method based on the sample-specific standard error for PLS regression, and compared with the univariate method endorsed by ISO 11483. The two methods gave similar results, both being effective for the intended purpose of estimating detection limits for PLS models. Although OSC and NAS allow isolating the analyte signal from the matrix signal, they provide no substantial improvement in terms of detection limits. The proposed method was used to the determine 2-ethylhexanol at concentrations from 20 to 1600ppm in an industrial ester. The detection limit obtained, round 100ppm, testifies to the ability of NIR spectroscopy to detect low concentrated analytes. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:318 / 323
页数:6
相关论文
共 24 条
[1]   Effect of orthogonal signal correction on the determination of compounds with very similar near infrared spectra [J].
Bertran, E ;
Iturriaga, H ;
Maspoch, S ;
Montoliu, I .
ANALYTICA CHIMICA ACTA, 2001, 431 (02) :303-311
[2]   NIR spectroscopy: a rapid-response analytical tool [J].
Blanco, M ;
Villarroya, I .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2002, 21 (04) :240-250
[3]  
BOOKSH KS, 1994, ANAL CHEM, V66, P15
[4]   Multivariate detection limits with fixed probabilities of error [J].
Boqué, R ;
Larrechi, MS ;
Rius, FX .
CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 1999, 45 (1-2) :397-408
[5]   DETECTION LIMITS WITH SPECIFIED ASSURANCE PROBABILITIES [J].
CLAYTON, CA ;
HINES, JW ;
ELKINS, PD .
ANALYTICAL CHEMISTRY, 1987, 59 (20) :2506-2514
[6]   Prediction error in least squares regression: Further critique on the deviation used in The Unscrambler [J].
Faber, K ;
Kowalski, BR .
CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 1996, 34 (02) :283-292
[7]   Standard error of prediction for multiway PLS 1. Background and a simulation study [J].
Faber, NM ;
Bro, R .
CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 2002, 61 (1-2) :133-149
[8]   Efficient computation of net analyte signal vector in inverse multivariate calibration models [J].
Faber, NM .
ANALYTICAL CHEMISTRY, 1998, 70 (23) :5108-5110
[9]  
Faber NM, 1998, J CHEMOMETR, V12, P405, DOI 10.1002/(SICI)1099-128X(199811/12)12:6<405::AID-CEM520>3.0.CO
[10]  
2-8