Evaluation of automated RNA-extraction technology and a qualitative HCV assay for sensitivity and detection of HCV RNA in pool-screening systems

被引:19
作者
Beld, M
Habibuw, MR
Rebers, SPH
Boom, R
Reesink, HW
机构
[1] Univ Amsterdam, Acad Med Ctr, Dept Med Microbiol, Sect Clin Virol, NL-1105 AZ Amsterdam, Netherlands
[2] Univ Amsterdam, Acad Med Ctr, Dept Gastroenterol & Hepatol, NL-1105 AZ Amsterdam, Netherlands
[3] Blood Bank N Holland Sanquin, Amsterdam, Netherlands
关键词
D O I
10.1046/j.1537-2995.2000.40050575.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND: The objective of this study was the evaluation of NAT technology for the detection of HCV RNA in plasma pools according to the recommendations of the Paul Ehrlich Institute (5000 IU/mL/donation) and the Committee for Proprietary Medical Products (100 IU/mL/manufacturing pool). STUDY DESIGN AND METHODS: Serial dilutions of both the EUROHEP standard (3,800 genome equivalents [geq]/mL; HCV genotype 1) and the World Health Organization (WHO) international standard (100,000 IU/mL; HCV genotype 1)were made in S/D plasma (ESPEP plasma, OctaPharma), which was nonreactive in serologic tests. Serial dilutions of plasma (2 mt) were used for extraction of HCV RNA with an automated version of a nucleic acid isolation method (NucliSens Extractor, Organon Teknika). HCV RNA was cc-extracted from 2 mt of plasma, together with 84 copies of an in vitro-synthesized single-strand RNA serving as internal extraction control (IC) to monitor the efficiency of extraction and PCR. Amplification and detection of both HCV RNA and IC RNA were performed with an automated PCR system and a qualitative HCV assay (COBAS Amplicor 2.0 HCV, Roche Diagnostics). RESULTS: A cutoff value of 16 geq per mt (10/10 runs [100% hit rate]) was found by using the EUROHEP standard, whereas the WHO international standard had a cutoff value of approximately 12 IU per mt (10/10 runs [100% hit rate]). The IC had a cutoff value of approximately 17.5 copies per mt (6/6 runs [100% hit rate]). Forty-two copies per mt of IC RNA were found in 282 of 284 runs (99% hit rate). The negative controls (ESDEP plasma) were negative in all experiments. Experiments with pool sizes of 12, 24, 48, and 96 using serial dilutions of the WHO international standard revealed a cutoff value of 8 IU per mt (100% hit rate). The EUROHEP standard and the WHO international standard were detected with a 50 percent detection endpoint of 5.2 geq per mt and 1.5 IU per mt, respectively. CONCLUSION: This test system (NucliSens Extractor, and the COBAS Amplicor 2.0 HCV assay) revealed a high sensitivity for HCV RNA; considering the proposed requirements for sensitivity of NAT assays for the detection of HCV RNA in donor plasma, pool sizes of about 400 donors are possible. These endpoint results indicated that 1 IU is equal to about 3.4 geq.
引用
收藏
页码:575 / 579
页数:5
相关论文
共 15 条
  • [1] TO C OR NOT TO C - THESE ARE THE QUESTIONS
    ALTER, HJ
    [J]. BLOOD, 1995, 85 (07) : 1681 - 1695
  • [2] Quantitative antibody responses to structural (core) and nonstructural (NS3, NS4, and NS5) hepatitis C virus proteins among seroconverting injecting drug users: Impact of epitope variation and relationship to detection of HCV RNA in blood
    Beld, M
    Penning, M
    van Putten, M
    Lukashov, V
    van den Hoek, A
    McMorrow, M
    Goudsmit, J
    [J]. HEPATOLOGY, 1999, 29 (04) : 1288 - 1298
  • [3] RAPID AND SIMPLE METHOD FOR PURIFICATION OF NUCLEIC-ACIDS
    BOOM, R
    SOL, CJA
    SALIMANS, MMM
    JANSEN, CL
    WERTHEIMVANDILLEN, PME
    VANDERNOORDAA, J
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1990, 28 (03) : 495 - 503
  • [4] SEQUENCE-ANALYSIS OF THE 5' NONCODING REGION OF HEPATITIS-C VIRUS
    BUKH, J
    PURCELL, RH
    MILLER, RH
    [J]. PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1992, 89 (11) : 4942 - 4946
  • [5] International collaborative study on the second EUROHEP HCV-RNA reference panel
    Damen, M
    Cuypers, HTM
    Zaaijer, HL
    Reesink, HW
    Schaasberg, WP
    Gerlich, WH
    Niesters, HGM
    Lelie, PN
    [J]. JOURNAL OF VIROLOGICAL METHODS, 1996, 58 (1-2) : 175 - 185
  • [6] DiDomenico N, 1996, CLIN CHEM, V42, P1915
  • [7] Nübling CM, 1998, INFUSIONSTHERAPIE, V25, P86
  • [8] ROGERS PM, 1996, EPFA NIBSC WORKSH NU, V72, P199
  • [9] Roggendorf M, 1996, J HEPATOL, V24, P26
  • [10] Collaborative study to assess the suitability of an HCV RNA reference sample for detection of an HCV RNA in plasma pools by PCR
    Saldanha, J
    Minor, P
    [J]. VOX SANGUINIS, 1996, 70 (03) : 148 - 151