Long-term experience with a preshaped left ventricular pacing lead

被引:8
作者
Ollitrault, J
Ritter, P
Mabo, P
Garrigue, S
Grossin, F
Lavergne, T
机构
[1] Hop Europeen Georges Pompidou, F-75908 Paris 15, France
[2] InParys, St Cloud, France
[3] CHU Pontchaillou, Rennes, France
[4] CHU Haut Eveque, Bordeaux, France
[5] ELA Med, Le Plessis Robinson, France
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2003年 / 26卷 / 01期
关键词
cardiac resynchronization therapy; left ventricular pacing; left ventricular pacing lead; congestive heart failure;
D O I
10.1046/j.1460-9592.2003.00013.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
This study describes a long-term experience with a new LV pacing lead. The study population consisted of 62 patients (85% men, 71 +/- 10 years old) with advanced dilated cardiomyopathy, in NYHA Class III or IV despite optimal drug therapy, and a QRS duration > 150 ms. Patients in sinus rhythm were implanted with a triple chamber pacemaker to maintain atrioventricular synchrony. A dual chamber pacemaker was implanted in patients in atria] fibrillation for biventricular pacing only. A clinical evaluation and interrogation of the resynchronization pacemaker were performed at implant, at 1 week (W1), one (M1), four (M4), and seven (W) months after implantation. A longer follow-up (2years) is available for patients implanted at the authors institution. LV measurements were pacing threshold at 0.5-ms pulse duration and pacing impedance. R wave amplitude (mV) was measured at the time of implantation only. The system was successfully implanted in 86% of patients with the latest design of the lead. Mean R wave amplitude at implant was 15 +/- 7 mV and mean pacing impedance was 1054 +/- 2540. Between implant (n = 38) and M7 (n = 15), pacing threshold rose from 0.73 +/- 0.54 to 1.57 +/- 0.60 V (P < 0.001). In conclusion, the situs lead was successfully implanted in a high percentage of patients. In addition, low pacing threshold and high impedance measured during follow-up are consistent with a low pacing current drain, ensuring a durable pulse generator longevity.
引用
收藏
页码:185 / 188
页数:4
相关论文
共 9 条
[1]  
ALONSO C, 2000, EUROPACE, V1, pD101
[2]  
BLANC JJ, 2000, EUROPACE, V1, pD101
[3]  
BLANC JJ, 2000, EUROPACE, V1, pD102
[4]   Effects of multisite biventricular pacing in patients with heart failure and intraventricular conduction delay. [J].
Cazeau, S ;
Leclercq, C ;
Lavergne, T ;
Walker, S ;
Varma, C ;
Linde, C ;
Garrigue, S ;
Kappenberger, L ;
Haywood, GA ;
Santini, M ;
Bailleul, C ;
Daubert, JC .
NEW ENGLAND JOURNAL OF MEDICINE, 2001, 344 (12) :873-880
[5]   Cardiac resynchronization therapy in advanced heart failure the multicenter InSync clinical study [J].
Gras, D ;
Leclercq, C ;
Tang, ASL ;
Bucknall, C ;
Luttikhuis, HO ;
Kirstein-Pedersen, A .
EUROPEAN JOURNAL OF HEART FAILURE, 2002, 4 (03) :311-320
[6]  
KUTARSKI A, 2000, EUROPACE, V1, pD307
[7]  
LECLERCQ C, 2000, EUROPACE, V1, pD103
[8]  
REX S, 2000, EUROPACE, V1, pD102
[9]  
SANTINI M, 2000, EUROPACE, V1, pD112