A 16-channel electrical stimulation system was implanted in a 39-year-old patient with T10 paraplegia to restore sit to stand, walking, and exercise functions. System implantation required two surgical sessions. in the first session, the posterior muscle set consisting of bilateral semimembranosus, adductor magnus, and gluteus maximus muscles were exposed and epimysial electrodes sutured at the point of greatest muscle contraction. Closed double helix intramuscular electrodes were implanted in the erector spinae. Two weeks later; epimysial electrodes were attached to the eight anterior muscles consisting of the tibialis anterior, sartorius, tensor fasciae latae, and vastus lateralis with all 16 electrode leads passed to the anterior abdominal wall. The electrodes were connected to two eight-channel stimulators placed in the iliac fossae, and the system was checked by activating the individual muscles. The implanted stimulators received stimulation instructions and power via a radio frequency link to an external control. Stimulation patterns for standing, walking, sitting, and exercise functions were chosen from a preprogrammed menu via a finger key pad. After 3 weeks of restricted patient activity all electrodes stimulated either the target muscle or had an acceptable spillover pattern. The patient is undergoing a 16-week rehabilitation course of stimulated exercises gradually increasing in intensity At the conclusion, the goal is to discharge the patient with the system for spontaneous use. Although long term followup is required to determine system reliability; preliminary clinical results indicate that targeted, repeatable, functional muscle contractions in the lower extremity can be achieved with a system consisting of epimysial electrodes.
引用
收藏
页码:236 / 242
页数:7
相关论文
共 22 条
[1]   RECRUITMENT PROPERTIES OF MONOPOLAR AND BIPOLAR EPIMYSIAL ELECTRODES [J].
GRANDJEAN, PA ;
MORTIMER, JT .
ANNALS OF BIOMEDICAL ENGINEERING, 1986, 14 (01) :53-66
[2]   DEVELOPMENT OF PERCUTANEOUS INTRAMUSCULAR ELECTRODE FOR MULTICHANNEL FES SYSTEM [J].
HANDA, Y ;
HOSHIMIYA, N ;
IGUCHI, Y ;
ODA, T .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1989, 36 (07) :705-710
[3]   REHABILITATION TECHNOLOGY FOR STANDING AND WALKING AFTER SPINAL-CORD INJURY [J].
JAEGER, RJ ;
YARKONY, GM ;
ROTH, EJ .
AMERICAN JOURNAL OF PHYSICAL MEDICINE & REHABILITATION, 1989, 68 (03) :128-133
[4]  
KAGAYA H, IN PRESS CLIN ORTHOP
[5]   IMPLANTABLE FUNCTIONAL NEUROMUSCULAR STIMULATION IN THE TETRAPLEGIC HAND [J].
KEITH, MW ;
PECKHAM, PH ;
THROPE, GB ;
STROH, KC ;
SMITH, B ;
BUCKETT, JR ;
KILGORE, KL ;
JATICH, JW .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 1989, 14A (03) :524-530
[6]  
KEITH MW, 1988, CLIN ORTHOP RELAT R, P25
[7]   ELECTRODE CHARACTERIZATION FOR FUNCTIONAL APPLICATION TO UPPER EXTREMITY FNS [J].
KILGORE, KL ;
PECKHAM, PH ;
KEITH, MW ;
THROPE, GB .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1990, 37 (01) :12-21
[8]   An implanted upper-extremity neuroprosthesis - Follow-up of five patients [J].
Kilgore, KL ;
Peckham, PH ;
Keith, MW ;
Thrope, GB ;
Wuolle, KS ;
Bryden, AM ;
Hart, RL .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (04) :533-541
[9]   CHRONICALLY IMPLANTED INTRAFASCICULAR RECORDING ELECTRODES [J].
LEFURGE, T ;
GOODALL, E ;
HORCH, K ;
STENSAAS, L ;
SCHOENBERG, A .
ANNALS OF BIOMEDICAL ENGINEERING, 1991, 19 (02) :197-207
[10]  
Marsolais E B, 1986, J Rehabil Res Dev, V23, P1