A new and highly sensitive continuous wave near infrared spectrophotometer with multiple detectors.

被引:52
作者
van der Sluijs, MC [1 ]
Colier, WNJM [1 ]
Houston, RJF [1 ]
Oeseburg, B [1 ]
机构
[1] Univ Nijmegen, Instrumentat Dept, NL-6500 HB Nijmegen, Netherlands
来源
PHOTON PROPAGATION IN TISSUES III, PROCEEDINGS OF | 1998年 / 3194卷
关键词
continuous wave near infrared spectrophotometry; instrumentation;
D O I
10.1117/12.301097
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
For a long time continuous wave near infrared instruments have been used to detect oxygenation changes in tissue. These instruments have proven to be reliable. The new generation of instruments, such as phase-modulated systems, or time-of-flight instruments, is not yet reliable enough for clinical applications. Most available: continuous wave near infrared instruments have low temporal resolution and low signal-to-noise ratio. For functional brain imaging, for example, a sensitive and fast instrument is needed. Therefore we developed the OXYMON, an instrument with a sample frequency up to 50 Hz and for optical densities up to 9 OD. The instrument uses 3 laser diodes, and is equipped with 1 or 2 avalanche photo detectors. Modular building techniques make maintenance easy.