The reliability of Microscan™ conventional and rapid panels to identify Staphylococcus aureus and detect methicillin resistance:: An evaluation using the tube coagulase test and mecA PCR

被引:7
作者
Farrell, DJ [1 ]
机构
[1] Queensland Hlth Pathol Serv, Toowoomba Lab, Dept Pathol, Microbiol Sect, Toowoomba, Qld 4350, Australia
关键词
Staphylococcus aureus; mecA; Microscan; Etest; PCR; methicillin resistance;
D O I
10.1080/00313029700169405
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
Microscan(TM) (Dade Diagnostics, Brisbane) Positive Combo Type 6 (312 panels) and Rapid Positive Breakpoint Type 1 (62 panels) were evaluated for Staphylococcus aureus identification, using the tube coagulase test (TC), and oxacillin susceptibility, using mecA. A total of 374 consecutive clinical staphylococci were tested, with TC and Microscan(TM) having 100% correlation (335 identified as S. aureus and 39 as coagulase negative staphylococci by both methods). A 93% correlation was observed between Microscan(TM) and mecA PCR for oxacillin susceptibility. No very major errors (0/374 false oxacillin susceptibility) and 26 (7%) major errors (26/374 false oxacillin resistance) were found showing false resistance to oxacillin to be a problem in our population. Oxacillin Etest (AB Biodisk, Sweden) was performed on all oxacillin resistant isolates. A bimodal distribution was observed between mecA positive and negative isolates. A testing algorithm (using the Microscan(TM) panels and Etest) was developed for this laboratory to detect mecA encoded methicillin resistance. Retrospective application of this algorithm to the 374 isolates gave 100% correlation with mecA detection.
引用
收藏
页码:406 / 410
页数:5
相关论文
共 26 条
[1]   METHICILLIN-RESISTANT STAPHYLOCOCCUS-AUREUS - OPTIMIZATION OF THE AGAR SCREEN PLATE METHOD [J].
ALFA, MJ ;
ONGSANSOY, E ;
HOBAN, S ;
MATHEW, T ;
GREENAWAY, C .
DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 1994, 18 (02) :121-124
[2]   IS431, A STAPHYLOCOCCAL INSERTION SEQUENCE-LIKE ELEMENT RELATED TO IS26 FROM PROTEUS-VULGARIS [J].
BARBERISMAINO, L ;
BERGERBACHI, B ;
WEBER, H ;
BECK, WD ;
KAYSER, FH .
GENE, 1987, 59 (01) :107-113
[3]  
BASCOMB S, 1991, PATHOL BIOL, V39, P466
[4]   LOW-LEVEL METHICILLIN RESISTANCE IN STRAINS OF STAPHYLOCOCCUS-AUREUS [J].
CHAMBERS, HF ;
ARCHER, G ;
MATSUHASHI, M .
ANTIMICROBIAL AGENTS AND CHEMOTHERAPY, 1989, 33 (04) :424-428
[5]  
COOMBS G, COMMUNICATION
[6]  
CRIST AE, 1992, GEN M AM SOC MICR
[7]   MULTIPLE MECHANISMS OF METHICILLIN RESISTANCE AND IMPROVED METHODS FOR DETECTION IN CLINICAL ISOLATES OF STAPHYLOCOCCUS-AUREUS [J].
DELENCASTRE, H ;
SAFIGUEIREDO, AM ;
URBAN, C ;
RAHAL, J ;
TOMASZ, A .
ANTIMICROBIAL AGENTS AND CHEMOTHERAPY, 1991, 35 (04) :632-639
[8]   EFFECT OF SOURCE OF MUELLER-HINTON AGAR ON DETECTION OF OXACILLIN RESISTANCE IN STAPHYLOCOCCUS-AUREUS USING A SCREENING METHODOLOGY [J].
HINDLER, JA ;
WARNER, NL .
JOURNAL OF CLINICAL MICROBIOLOGY, 1987, 25 (04) :734-735
[9]  
JEVONS MP, 1961, BRIT MED J, V1, P124, DOI 10.1136/bmj.1.5219.124-a
[10]   IDENTIFICATION OF STAPHYLOCOCCUS SPECIES AND SUBSPECIES WITH THE MICROSCAN POS ID AND RAPID POS ID PANEL SYSTEMS [J].
KLOOS, WE ;
GEORGE, CG .
JOURNAL OF CLINICAL MICROBIOLOGY, 1991, 29 (04) :738-744