A single-channel implantable microstimulator for functional neuromuscular stimulation

被引:90
作者
Ziaie, B
Nardin, MD
Coghlan, AR
Najafi, K
机构
[1] Center for Integrated Sensors and Circuits, Department of Electrical Engineering and Computer Science, University of Michigan, Ann Arbor MI 48109-2122
关键词
biotelemetry; functional electrical stimulation; implantable devices; medical microsystems; rehabilitative devices;
D O I
10.1109/10.634643
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper describes a single-channel implantable microstimulator for functional neuromuscular stimulation, This device measures 2x2x10 mm(3) and can be inserted into paralyzed muscle groups by expulsion from a hypodermic needle, Power and data to the device are supplied from outside by RF telemetry using an amplitude-modulated 2-MHz RF carrier generated using a high-efficiency class-E transmitter, The transmitted signal carries a 5-b address which selects one of the 32 possible microstimulators. The selected device then delivers up to 2 mu C of charge stored in a tantalum chip capacitor for up to 200 mu s (10 mA) into loads of <800 Omega through a high-current thin-film iridium-oxide (IrOx) electrode (similar to 0.3 mm(2) in area), A bi-CMOS receiver circuitry is used to: generate two regulated voltage supplies (4.5 and 9 V), recover a 2-MHz clock from the carrier, demodulate the address code, and activate the output current delivery circuitry upon the reception of an external command, The overall power dissipation of the receiver circuitry is 45-55 mW. The implant is hermetically packaged using a custom-made glass capsule.
引用
收藏
页码:909 / 920
页数:12
相关论文
共 34 条
  • [1] Adams A, 1988, VLSI TECHNOLOGY
  • [2] AKIN T, 1990, IEEE SOL STAT SENS A, P145
  • [3] BRINDLY G, 1963, J PHYS, V195, P479
  • [4] A FLEXIBLE, PORTABLE SYSTEM FOR NEUROMUSCULAR STIMULATION IN THE PARALYZED UPPER EXTREMITY
    BUCKETT, JR
    PECKHAM, PH
    THROPE, GB
    BRASWELL, SD
    KEITH, MW
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (11) : 897 - 904
  • [5] Electromagnetic field effects and mechanisms - In search of an anchor
    Foster, KR
    [J]. IEEE ENGINEERING IN MEDICINE AND BIOLOGY MAGAZINE, 1996, 15 (04): : 50 - 56
  • [6] Gray P. R., 1993, ANAL DESIGN ANALOG I
  • [7] THERAPEUTIC APPLICATIONS OF ELECTROMAGNETIC POWER
    GUY, AW
    LEHMANN, JF
    STONEBRIDGE, JB
    [J]. PROCEEDINGS OF THE IEEE, 1974, 62 (01) : 55 - 75
  • [8] HAMBRECHT FT, 1977, FUNCTIONAL ELECTRICA
  • [9] Haus H.A., 1989, ELECTROMAGNETIC FIEL
  • [10] AN IMPLANTABLE CMOS CIRCUIT INTERFACE FOR MULTIPLEXED MICROELECTRODE RECORDING ARRAYS
    JI, J
    WISE, KD
    [J]. IEEE JOURNAL OF SOLID-STATE CIRCUITS, 1992, 27 (03) : 433 - 443