Temperature-controlled irrigated tip radiofrequency catheter ablation: Comparison of in vivo and in vitro lesion dimensions for standard catheter and irrigated tip catheter with minimal infusion rate

被引:52
作者
Petersen, HH [1 ]
Chen, X [1 ]
Pietersen, A [1 ]
Svendsen, JH [1 ]
Haunso, S [1 ]
机构
[1] Univ Copenhagen, Natl Hosp, Med Dept B2142, DK-2100 Copenhagen O, Denmark
关键词
catheter ablation; ventricular arrhythmias; radiofrequency ablation; cooled tip electrode;
D O I
10.1111/j.1540-8167.1998.tb00928.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: In patients with ventricular tachycardias due to structural heart disease, catheter ablation cures < 60% partly due to the limited lesion size after conventional radiofrequency ablation, Irrigated tip radiofrequency ablation using power control and high infusion rates enlarges lesion size, but has increased risk of cratering, The present study explores irrigated tip catheter ablation in temperature-controlled mode, target temperature 60 degrees C, using an irrigation rate of 1 mL/min, comparing this to conventional catheter technique, target temperature 80 degrees C. Methods and Results: In vivo anesthetized pigs were ablated in the left ventricle, In vitro strips of porcine left ventricular myocardium were ablated in a tissue bath. Lesion volume was significantly larger after irrigated tip ablation both in vivo (544 +/- 218 vs 325 +/- 194 mm(3), P < 0.01) and in vitro (286 +/- 113 vs 179 +/- 23 mm(3), P < 0.001), The incidence of cratering was not significantly different between the two groups. In vivo, no coagulum formation on part of the catheter tip was seen after irrigated tip ablation as opposed to 52% of the applications with conventional ablation (P < 0.05). Conclusion: We conclude that temperature-controlled radiofrequency ablation with irrigated tip catheters using low target temperature and low infusion rate enlarges lesion size without increasing the incidence of cratering and reduces coagulum formation of the tip.
引用
收藏
页码:409 / 414
页数:6
相关论文
共 26 条
[1]  
ALEXANDER ME, 1997, PHYS MED BIOL, V20, P1087
[2]   EXPERIMENTAL-VERIFICATION OF BIOHEAT TRANSFER THEORIES - MEASUREMENT OF TEMPERATURE PROFILES AROUND LARGE ARTIFICIAL VESSELS IN PERFUSED TISSUE [J].
CREZEE, J ;
LAGENDIJK, JJW .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (07) :905-923
[3]   CATHETER ABLATION OF VENTRICULAR-TACHYCARDIA [J].
DAOUD, E ;
MORADY, F .
CURRENT OPINION IN CARDIOLOGY, 1995, 10 (01) :21-25
[4]  
DEMAZUMDER D, 1997, PACE, V20, P1203
[5]  
DEMAZUMDER D, 1997, PACE, V20, P1076
[6]  
Dorwarth U., 1996, European Heart Journal, V17, P270
[7]   TISSUE HEATING DURING RADIOFREQUENCY CATHETER ABLATION - A THERMODYNAMIC MODEL AND OBSERVATIONS IN ISOLATED PERFUSED AND SUPERFUSED CANINE RIGHT VENTRICULAR FREE WALL [J].
HAINES, DE ;
WATSON, DD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (06) :962-976
[8]  
HAINES DE, 1994, CATHETER ABLATION AR, P105
[9]  
HAINES DE, 1995, CARDIAC ELECTROPHYSI, P1442
[10]   COAGULATION OF VENTRICULAR MYOCARDIUM USING RADIOFREQUENCY ALTERNATING-CURRENT - BIO-PHYSICAL ASPECTS AND EXPERIMENTAL FINDINGS [J].
HAVERKAMP, W ;
HINDRICKS, G ;
GULKER, H ;
RISSEL, U ;
PFENNINGS, W ;
BORGGREFE, M ;
BREITHARDT, G .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (01) :187-195