Comparison of 3-D maximum a posteriori and filtered backprojection algorithms for high-resolution animal imaging with microPET

被引:103
作者
Chatziioannou, A [1 ]
Qi, J
Moore, A
Annala, A
Nguyen, K
Leahy, R
Cherry, SR
机构
[1] Univ Calif Los Angeles, Sch Med, Dept Mol & Med Pharmacol, Crump Inst Biol Imaging, Los Angeles, CA 90095 USA
[2] Univ Calif Berkeley, Ctr Funct Imaging, Berkeley, CA 94720 USA
[3] Univ So Calif, Inst Signal & Image Proc, Los Angeles, CA 90089 USA
关键词
image reconstruction; positron emission tomography; small animal imaging;
D O I
10.1109/42.870260
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
We have evaluated the performance of two three-dimensional (3-D) reconstruction algorithms with data acquired from microPET, a high resolution tomograph dedicated to small animal imaging. The first was a linear filtered-backprojection algorithm (FBP) with reprojection of the missing data, and the second was a statistical maximum a posteriori probability algorithm (MAP), The two algorithms were evaluated in terms of their resolution performance, both in phantoms and in vivo. Sixty independent realizations of a phantom simulating the brain of a baby monkey were acquired, each containing three million counts. Each of these realizations was reconstructed independently with both algorithms. The ensemble of the 60 reconstructed realizations was used to estimate the standard deviation as a measure of the noise for each reconstruction algorithm, More detail was recovered in the MAP reconstruction without an increase in noise relative to FBP. Studies in a simple cylindrical compartment phantom demonstrated improved recovery of known activity ratios with MAP. Finally, in vivo studies also demonstrated a clear improvement in spatial resolution using the MAP algorithm. The quantitative accuracy of the MAP reconstruction was also evaluated by comparison with autoradiography and direct well counting of tissue samples and was shown to be superior.
引用
收藏
页码:507 / 512
页数:6
相关论文
共 18 条
[1]  
Chatziioannou AF, 1999, J NUCL MED, V40, P1164
[2]   PARALLELIZATION OF THE EM ALGORITHM FOR 3-D PET IMAGE-RECONSTRUCTION [J].
CHEN, CM ;
LEE, SY ;
CHO, ZH .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1991, 10 (04) :513-522
[3]   MicroPET: A high resolution PET scanner for imaging small animals [J].
Cherry, SR ;
Shao, Y ;
Silverman, RW ;
Meadors, K ;
Siegel, S ;
Chatziioannou, A ;
Young, JW ;
Jones, WF ;
Moyers, JC ;
Newport, D ;
Boutefnouchet, A ;
Farquhar, TH ;
Andreaco, M ;
Paulus, MJ ;
Binkley, DM ;
Nutt, R ;
Phelps, ME .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :1161-1166
[4]   Exact and approximate rebinning algorithms for 3-D PET data [J].
Defrise, M ;
Kinahan, PE ;
Townsend, DW ;
Michel, C ;
Sibomana, M ;
Newport, DF .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (02) :145-158
[5]   An evaluation of exact and approximate 3-D reconstruction algorithms for a high-resolution, small-animal PET scanner [J].
Farquhar, TH ;
Chatziioannou, A ;
Cherry, SR .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (06) :1073-1080
[6]   Spatial resolution properties of penalized-likelihood image reconstruction: Space-invariant tomographs [J].
Fessler, JA ;
Rogers, WL .
IEEE TRANSACTIONS ON IMAGE PROCESSING, 1996, 5 (09) :1346-1358
[7]   ASSESSMENT OF ACCURACY OF PET UTILIZING A 3-D PHANTOM TO SIMULATE THE ACTIVITY DISTRIBUTION OF [18F]FLUORODEOXYGLUCOSE UPTAKE IN THE HUMAN BRAIN [J].
HOFFMAN, EJ ;
CUTLER, PD ;
GUERRERO, TM ;
DIGBY, WM ;
MAZZIOTTA, JC .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1991, 11 (02) :A17-A25
[8]   A SYSTEM FOR THE 3D RECONSTRUCTION OF RETRACTED-SEPTA PET DATA USING THE EM ALGORITHM [J].
JOHNSON, CA ;
YAN, YC ;
CARSON, RE ;
MARTINO, RL ;
DAUBEWITHERSPOON, ME .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) :1223-1227
[9]   ANALYTIC 3D IMAGE-RECONSTRUCTION USING ALL DETECTED EVENTS [J].
KINAHAN, PE ;
ROGERS, JG .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1989, 36 (01) :964-968
[10]  
Myers R., 1996, QUANTIFICATION BRAIN, P12