An automated assay for measuring serum ascorbic acid with use of 4-hydroxy-2,2,6,6-tetramethylpiperidinyloxy, free radical and o-phenylenediamine

被引:22
作者
Ihara, H
Matsumoto, N
Shino, Y
Aoki, Y
Hashizume, N
Nanba, S
Urayama, T
机构
[1] Toho Univ, Sch Med, Ohashi Hosp, Dept Lab Med,Meguro Ku, Tokyo 1538515, Japan
[2] Toho Univ, Sch Med, Omori Hosp, Dept Lab Med,Ota Ku, Tokyo 1438541, Japan
关键词
vitamin C determination; ascorbic acid; automated assay; dehydroascorbic acid; vitamin C deficiency; o-phenylenediamine; 4-hydroxy-2,2,6,6-tetramethylpiperidinyloxy free; radical;
D O I
10.1016/S0009-8981(00)00356-9
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 [基础医学];
摘要
We developed a novel, cost-effective, and automated assay for ascorbic acid (AsA) in serum using a COBAS MIRA S analyzer (Roche Diagnostic System). Our method has a wide dynamic range and covers AsA concentrations from well below the lower reference interval to well above it. AsA is oxidized by 4-hydroxy-2,2,6,6-tetramethylpiperidinyloxy, free radical (TEMPO) to dehydroascorbic acid (DAsA). The latter condenses with o-phenylenediamine (OPDA) to form a quinoxaline derivative that absorbs light at 340 nm. The change in absorbance at 340 nm is proportional to the concentration of AsA in the specimen. The automated system permitted the assay of 65 specimens per hour at a cost of approximately US$ 0.01 per specimen for reagents. The assay can be applied directly to serum specimens (direct method) and also to sera with a prior deproteinization step with metaphosphoric acid. The detection limit for the direct serum assays is 0.8 vs. 0.4 mg/l with the deproteinization method. The recovery of AsA from a supplemented serum peril was of >95% for both procedures. We used four distinct methods on 66 patients sera. The direct method for AsA correlated well with an HPLC method (r = 0.964, P < 0.001); the direct method also correlated well with a method that uses AsA oxidase (r = 0.975. P < 0.001). The deproteinization method correlated well with HPLC (r = 0.981, P < 0.001), and with the AsA oxidase procedure (r = 0.994, P < 0.001). Ten within-day determinations on a serum pool gave a C.V. < 4.3% for both the direct and deproteinization procedures. The between-day assays of the same serum pool over 10 days gave a C.V. of <6.7% by both methods. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:193 / 204
页数:12
相关论文
共 15 条
[1]
BUDAVARI S, 1996, MERCK INDEX, P139
[2]
HEMOGLOBIN DETERMINATION IN PLASMA OR SERUM BY 1ST-DERIVATIVE RECORDING SPECTROPHOTOMETRY - EVALUATION OF THE PROCEDURE OF SOLONI, CUNNINGHAM, AND AMAZON [J].
COPELAND, BE ;
DYER, PJ ;
PESCE, AJ .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1989, 92 (05) :619-624
[3]
DOUMAS BT, 1985, CLIN CHEM, V31, P1779
[4]
HORNING D, 1982, J JPN SOC CLIN NUTR, V3, P127
[5]
Ihara H, 1999, Nihon Rinsho, V57 Suppl, P147
[6]
IHARA H, IN PRESS CLIN CHEM E
[7]
IHARA H, 1999, J ANAL BIOSCI, V22, P283
[8]
JACOB RA, 1995, CLIN GUIDE LAB TESTS, P640
[9]
JACOBS DS, 1996, BLOOD LAB TESTS HDB, P84
[10]
MCCORMICK DB, 1994, TIETZ TXB CLIN CHEM, P1311