SmartHand tactile display: A new concept for providing sensory feedback in hand prostheses

被引:26
作者
Antfolk, Christian [1 ]
Balkenius, Christian [2 ]
Rosen, Birgitta [3 ]
Lundborg, Goran [3 ]
Sebelius, Fredrik [1 ]
机构
[1] Lund Univ, Dept Elect Measurements, SE-22100 Lund, Sweden
[2] Lund Univ Cognit Sci, Lund, Sweden
[3] Malmo Univ Hosp, Dept Hand Surg, Malmo, Sweden
关键词
Sensory feedback; hand prosthesis; haptic perception; SUBSTITUTION; SENSATION; AMPUTEES; SKIN;
D O I
10.3109/02844310903259090
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
A major drawback with myoelectric prostheses is that they do not provide the user with sensory feedback. Using a new principle for sensory feedback, we did a series of experiments involving 11 healthy subjects. The skin on the volar aspect of the forearm was used as the target area for sensory input. Experiments included discrimination of site of stimuli and pressure levels at a single stimulation point. A tactile display based on digital servomotors with one actuating element for each of the five fingers was used as a stimulator on the forearm. The results show that the participants were able to discriminate between three fingers with an accuracy of 97%, between five fingers with an accuracy of 82%, and between five levels with an accuracy of 79%. The tactile display may prove a helpful tool in providing amputees with sensory feedback from a prosthetic hand by transferring tactile stimuli from the prosthetic hand to the skin at forearm level.
引用
收藏
页码:50 / 53
页数:4
相关论文
共 17 条
[1]  
[Anonymous], 2007, JPO, Journal of Prosthetics and Orthotics, DOI 10.1097/JPO.0b013e3180459df6
[2]  
Billock J N., 1995, Rehabilitation of the Hand: Surgery and Therapy, P1189
[3]   On the shared control of an EMG-controlled prosthetic hand: Analysis of user-prosthesis interaction [J].
Cipriani, Christian ;
Zaccone, Franco ;
Micera, Silvestro ;
Carrozza, M. Chiara .
IEEE TRANSACTIONS ON ROBOTICS, 2008, 24 (01) :170-184
[4]   Upper limb amputees can be induced to experience a rubber hand as their own [J].
Ehrsson, H. Henrik ;
Rosen, Birgitta ;
Stockselius, Anita ;
Ragno, Christina ;
Kohler, Peter ;
Lundborg, Goran .
BRAIN, 2008, 131 :3443-3452
[5]   Perceptual phenomena after unilateral arm amputation:: a pre-post-surgical comparison [J].
Grüsser, SM ;
Winter, C ;
Schaefer, M ;
Fritzsche, K ;
Benhidjeb, T ;
Tunn, PU ;
Schlag, PM ;
Flor, H .
NEUROSCIENCE LETTERS, 2001, 302 (01) :13-16
[6]  
Arieta AH, 2005, P ANN INT IEEE EMBS, P6919
[7]  
Hubbard S., 1995, Rehabilitation of the Hand: Surgery and Therapy, P1241
[8]   Assessment of sensibility after nerve injury and repair: A systematic review of evidence for validity, reliability and responsiveness of tests [J].
Jerosch-Herold, C .
JOURNAL OF HAND SURGERY-BRITISH AND EUROPEAN VOLUME, 2005, 30B (03) :252-264
[9]  
JOHANSSON RS, 1984, EXP BRAIN RES, V56, P550
[10]   ELECTROTACTILE AND VIBROTACTILE DISPLAYS FOR SENSORY SUBSTITUTION SYSTEMS [J].
KACZMAREK, KA ;
WEBSTER, JG ;
BACHYRITA, P ;
TOMPKINS, WJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1991, 38 (01) :1-16