Real-time immunoassay of ferritin using surface plasmon resonance biosensor

被引:47
作者
Cui, XQ [1 ]
Yang, F [1 ]
Sha, YF [1 ]
Yang, XR [1 ]
机构
[1] Chinese Acad Sci, Changchun Inst Appl Chem, State Key Lab Electroanalyt Chem, Changchun 130022, Jilin, Peoples R China
基金
中国国家自然科学基金;
关键词
surface plasmon resonance; immunoassay; ferritin; sandwich enhancement;
D O I
10.1016/S0039-9140(03)00043-2
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A surface plasmon resonance (SPR) biosensor was used for the first time to determine the concentration of ferritin in both HBS-EP buffer and serum. The monoclonal antibody was immobilized on the carboxymethyl dextran-modified gold surface by an amine coupling method. The interaction of antibody with antigen was monitored in real-time. The signal was enhanced by sandwich amplification strategy to improve the sensitivity and specificity of the immunoassay, especially in serum. The linear range of the assay in serum is over 30-200 ng ml with the detection limit of 28 ng ml(-1). The sensitivity, specificity, and reproducibility of the assay are satisfactory. The analyte and enhancement antibody-binding surface could be regenerated by pH 2.0 glycine-HCl buffer and the same antibody-immobilized surface could be used for more than 50 cycles of ferritin binding and regeneration. (C) 2003 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:53 / 61
页数:9
相关论文
共 21 条
  • [1] Development of an immunosensor for human ferritin, a nonspecific tumor marker, based on a quartz crystal microbalance
    Chou, SF
    Hsu, WL
    Hwang, JM
    Chen, CY
    [J]. ANALYTICA CHIMICA ACTA, 2002, 453 (02) : 181 - 189
  • [2] BIOSPECIFIC INTERACTION ANALYSIS USING SURFACE-PLASMON RESONANCE DETECTION APPLIED TO KINETIC, BINDING-SITE AND CONCENTRATION ANALYSIS
    FAGERSTAM, LG
    FROSTELLKARLSSON, A
    KARLSSON, R
    PERSSON, B
    RONNBERG, I
    [J]. JOURNAL OF CHROMATOGRAPHY, 1992, 597 (1-2): : 397 - 410
  • [3] Fairbanks VF., 1994, TIETZ TXB CLIN CHEM, P1974
  • [4] Evaluation of a particle-enhanced turbidimetric immunoassay for the measurement of ferritin:: Application to patients participating in an autologous blood transfusion program
    Gómez, F
    Simó, JM
    Camps, J
    Clivillé, X
    Bertran, N
    Ferré, N
    Bofill, C
    Joven, J
    [J]. CLINICAL BIOCHEMISTRY, 2000, 33 (03) : 191 - 196
  • [5] SOLID-PHASE RADIOIMMUNOASSAY FOR SERUM FERRITIN
    HALLIDAY, JW
    GERA, KL
    POWELL, LW
    [J]. CLINICA CHIMICA ACTA, 1975, 58 (03) : 207 - 214
  • [6] IMMOBILIZATION OF PROTEINS TO A CARBOXYMETHYLDEXTRAN-MODIFIED GOLD SURFACE FOR BIOSPECIFIC INTERACTION ANALYSIS IN SURFACE-PLASMON RESONANCE SENSORS
    JOHNSSON, B
    LOFAS, S
    LINDQUIST, G
    [J]. ANALYTICAL BIOCHEMISTRY, 1991, 198 (02) : 268 - 277
  • [7] Efficacy of oral iron supplementation is not enhanced by additional intravenous iron during autologous blood donation
    Kasper, SM
    Lazansky, H
    Stark, C
    Klimek, M
    Laubinger, R
    Borner, U
    [J]. TRANSFUSION, 1998, 38 (08) : 764 - 770
  • [8] LEVI S, 1988, J BIOL CHEM, V263, P18086
  • [9] A NOVEL HYDROGEL MATRIX ON GOLD SURFACES IN SURFACE-PLASMON RESONANCE SENSORS FOR FAST AND EFFICIENT COVALENT IMMOBILIZATION OF LIGANDS
    LOFAS, S
    JOHNSSON, B
    [J]. JOURNAL OF THE CHEMICAL SOCIETY-CHEMICAL COMMUNICATIONS, 1990, (21) : 1526 - 1528
  • [10] BIOANALYSIS WITH SURFACE-PLASMON RESONANCE
    LOFAS, S
    MALMQVIST, M
    RONNBERG, I
    STENBERG, E
    LIEDBERG, B
    LUNDSTROM, I
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 1991, 5 (1-4) : 79 - 84