Finite volume analysis of temperature effects induced by active MRI implants with cylindrical symmetry: 1. Properly working devices

被引:6
作者
Busch, Martin H. J. [1 ,4 ]
Vollmann, Wolfgang [2 ]
Schnorr, Joerg [3 ]
Groenemeyer, Dietrich H. W. [4 ]
机构
[1] Res & Dev Ctr Microtherapy EFMT, D-44799 Bochum, Germany
[2] TFH Univ Appl Sci, D-13353 Berlin, Germany
[3] Humboldt Univ, Med Fak, Inst Radiol, Charite, D-10117 Berlin, Germany
[4] Univ Witten Herdecke, Gronemeyer Inst Microtherapy, D-44799 Bochum, Germany
关键词
Quality Factor; Power Loss; Specific Absorption Rate; Simulation Volume; Faraday Cage;
D O I
10.1186/1475-925X-4-25
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Active Magnetic Resonance Imaging implants are constructed as resonators tuned to the Larmor frequency of a magnetic resonance system with a specific field strength. The resonating circuit may be embedded into or added to the normal metallic implant structure. The resonators build inductively coupled wireless transmit and receive coils and can amplify the signal, normally decreased by eddy currents, inside metallic structures without affecting the rest of the spin ensemble. During magnetic resonance imaging the resonators generate heat, which is additional to the usual one described by the specific absorption rate. This induces temperature increases of the tissue around the circuit paths and inside the lumen of an active implant and may negatively influence patient safety. Methods: This investigation provides an overview of the supplementary power absorbed by active implants with a cylindrical geometry, corresponding to vessel implants such as stents, stent grafts or vena cava filters. The knowledge of the overall absorbed power is used in a finite volume analysis to estimate temperature maps around different implant structures inside homogeneous tissue under worst-case assumptions. The "worst-case scenario" assumes thermal heat conduction without blood perfusion inside the tissue around the implant and mostly without any cooling due to blood flow inside vessels. Results: The additional power loss of a resonator is proportional to the volume and the quality factor, as well as the field strength of the MRI system and the specific absorption rate of the applied sequence. For properly working devices the finite volume analysis showed only tolerable heating during MRI investigations in most cases. Only resonators transforming a few hundred mW into heat may reach temperature increases over 5 K. This requires resonators with volumes of several ten cubic centimeters, short inductor circuit paths with only a few 10 cm and a quality factor above ten. Using MR sequences, for which the MRI system manufacturer declares the highest specific absorption rate of 4 W/kg, vascular implants with a realistic construction, size and quality factor do not show temperature increases over a critical value of 5 K. Conclusion: The results show dangerous heating for the assumed "worst-case scenario" only for constructions not acceptable for vascular implants. Realistic devices are safe with respect to temperature increases. However, this investigation discusses only properly working devices. Ruptures or partial ruptures of the wires carrying the electric current of the resonance circuits or other defects can set up a power source inside an extremely small volume. The temperature maps around such possible "hot spots" should be analyzed in an additional investigation.
引用
收藏
页数:17
相关论文
共 25 条
[1]  
ATHEY T W, 1982, Magnetic Resonance Imaging, V1, P149, DOI 10.1016/0730-725X(82)90206-5
[2]   MR imaging of vascular stents: effects of susceptibility, flow, and radiofrequency eddy currents [J].
Bartels, LW ;
Smits, HFM ;
Bakker, CJG ;
Viergever, MA .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2001, 12 (03) :365-371
[3]   MAXIMIZING CONTRAST TO NOISE WITH INDUCTIVELY COUPLED IMPLANTED COILS [J].
FARMER, THR ;
COFER, GP ;
JOHNSON, GA .
INVESTIGATIVE RADIOLOGY, 1990, 25 (05) :552-558
[4]   SPECIFIC RESISTANCE OF BIOLOGICAL MATERIAL-A COMPENDUM OF DATA FOR BIOMEDICAL ENGINEER AND PHYSIOLOGIST [J].
GEDDES, LA ;
BAKER, LE .
MEDICAL & BIOLOGICAL ENGINEERING, 1967, 5 (03) :271-&
[5]   Transport lattice models of heat transport in skin with spatially heterogeneous, temperature-dependent perfusion [J].
Gowrishankar, T. R. ;
Stewart, Donald A. ;
Martin, Gregory T. ;
Weaver, James C. .
BIOMEDICAL ENGINEERING ONLINE, 2004, 3 (1)
[6]  
Hand JW, 1999, MAGN RESON MED, V42, P183, DOI 10.1002/(SICI)1522-2594(199907)42:1<183::AID-MRM24>3.0.CO
[7]  
2-I
[8]   Effects of hyperthermia on the peripheral nervous system: a review [J].
Haveman, J ;
Van der Zee, J ;
Wondergem, J ;
Hoogeveen, JF ;
Hulshof, MCCM .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2004, 20 (04) :371-391
[9]   Coronary arterial stents: Safety and artifacts during MR imaging [J].
Hug, J ;
Nagel, E ;
Bornstedt, A ;
Schnackenburg, B ;
Oswald, H ;
Fleck, E .
RADIOLOGY, 2000, 216 (03) :781-787
[10]  
Johnson C., 1972, P IEEE, V60