Microwave, irrigated, pulsed, or conventional radiofrequency energy source: Which energy source for which catheter ablation?

被引:11
作者
Erdogan, A
Grumbrecht, S
Neumann, T
Neuzner, J
Pitschner, HF
机构
[1] Univ Giessen, Dept Cardiol Angiol, D-35390 Giessen, Germany
[2] Kerckhoff Clin, Dept Cardiol, Bad Nauheim, Germany
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2003年 / 26卷 / 01期
关键词
radiofrequency catheter ablation; microwave; irrigated catheter; pulsed energy delivery;
D O I
10.1046/j.1460-9592.2003.00082.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The aim of the study was to compare the diameter of endomyocardial lesions induced with the delivery of microwave, cooled, or pulsed energy versus conventional RF energy. In vitro tests were performed in fresh endomyocardial preparations of pig hearts in a 10-L bath of NaCl 0.9% solution at 37degreesC and constant 1.5 L/min flow. Ablation 7 Er catheters with 4-mm tip electrodes were used, except for the delivery of microwave energy. Energy delivery time was set to 60 s/50 W in all experiments. Cooled energy delivery was performed with a closed irrigation catheter. Pulsed energy delivery was performed using a special controller with a duty-cycle of 5 ms. Microwave energy was delivered with a 2.5-GHz generator and 10-mm antenna. Electrode temperature and impedance were measured simultaneously. After ablation, lesion length, width, and depth were measured with microcalipers, and volume calculated by a formula for ellipsoid bodies. Each energy delivery mode was tested in ten experiments. The deepest lesions were created with cooled energy delivery and the largest volume by microwave energy delivery. Pulsed RF produced significantly deeper lesions than conventional RF energy delivery. Cooled or pulsed RF energy delivery created deeper transmural lesions than conventional RE To create linear lesions at anatomically complex sites (isthmus), microwave energy seemed superior by rapidly creating deep and long lesions.
引用
收藏
页码:504 / 506
页数:3
相关论文
共 11 条
[1]   RADIOFREQUENCY CATHETER ABLATION OF ACCESSORY ATRIOVENTRICULAR CONNECTIONS IN 250 PATIENTS - ABBREVIATED THERAPEUTIC APPROACH TO WOLFF-PARKINSON-WHITE SYNDROME [J].
CALKINS, H ;
LANGBERG, J ;
SOUSA, J ;
ELATASSI, R ;
LEON, A ;
KOU, W ;
KALBFLEISCH, S ;
MORADY, F .
CIRCULATION, 1992, 85 (04) :1337-1346
[2]   Effects of linear, irrigated-tip radiofrequency ablation in porcine healed anterior infarction [J].
Callans, DJ ;
Ren, JF ;
Narula, N ;
Michele, J ;
Marchlinski, FE ;
Dillon, SM .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2001, 12 (09) :1037-1042
[3]   RADIOFREQUENCY ABLATION OF THE INFERIOR VENA-CAVA - TRICUSPID-VALVE ISTHMUS IN COMMON ATRIAL-FLUTTER [J].
COSIO, FG ;
LOPEZGIL, M ;
GOICOLEA, A ;
ARRIBAS, F ;
BARROSO, JL .
AMERICAN JOURNAL OF CARDIOLOGY, 1993, 71 (08) :705-709
[4]  
Erdogan A, 2000, PACE, V23, P1852
[5]   Right and left atrial radiofrequency catheter therapy of paroxysmal atrial fibrillation [J].
Haissaguerre, M ;
Jais, P ;
Shah, DC ;
Gencel, L ;
Pradeau, V ;
Garrigues, S ;
Chouairi, S ;
Hocini, M ;
LeMetayer, P ;
Roudaut, R ;
Clementy, J .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1996, 7 (12) :1132-1144
[6]   TREATMENT OF SUPRAVENTRICULAR TACHYCARDIA DUE TO ATRIOVENTRICULAR NODAL REENTRY BY RADIOFREQUENCY CATHETER ABLATION OF SLOW-PATHWAY CONDUCTION [J].
JACKMAN, WM ;
BECKMAN, KJ ;
MCCLELLAND, JH ;
WANG, XZ ;
FRIDAY, KJ ;
ROMAN, CA ;
MOULTON, KP ;
TWIDALE, N ;
HAZLITT, HA ;
PRIOR, MI ;
OREN, J ;
OVERHOLT, ED ;
LAZZARA, R .
NEW ENGLAND JOURNAL OF MEDICINE, 1992, 327 (05) :313-318
[7]   Irrigated radiofrequency catheter ablation [J].
Keane, D .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2001, 12 (09) :1043-1045
[8]   RADIOFREQUENCY CATHETER ABLATION - THE EFFECT OF ELECTRODE SIZE ON LESION VOLUME INVIVO [J].
LANGBERG, JJ ;
LEE, MA ;
CHIN, MC ;
ROSENQVIST, M .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (10) :1242-1248
[9]   BIOPHYSICS AND PATHOLOGY OF CATHETER ENERGY DELIVERY SYSTEMS [J].
NATH, S ;
HAINES, DE .
PROGRESS IN CARDIOVASCULAR DISEASES, 1995, 37 (04) :185-204
[10]   Linear ablation using an irrigated electrode - Electrophysiologic and histologic lesion evolution comparison with ablation utilizing a non-irrigated electrode [J].
Schwartzman, D ;
Parizhskaya, M ;
Devine, WA .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2001, 5 (01) :17-26