Biomechanical evaluation of the modified double-plating fixation for the distal radius fracture

被引:48
作者
Cheng, Hsin-Yi Kathy
Lin, Chun-Li
Lin, Yu-Hao
Chen, Alvin Chao-Yu
机构
[1] Chang Gung Univ, Dept Mech Engn, Tao Yuan 333, Taiwan
[2] Chang Gung Univ, Grad Inst Mech Engn, Tao Yuan 333, Taiwan
[3] Chang Gung Univ, Dept Phys Therapy, Tao Yuan 333, Taiwan
[4] Chang Gung Univ, Grad Inst Med Mechatron, Tao Yuan 333, Taiwan
[5] Chang Gung Mem Hosp, Dept Orthopaed Surg, Tao Yuan 333, Taiwan
关键词
distal radius fracture; fracture fixation; double-plating technique; finite element; biomechanical analysis;
D O I
10.1016/j.clinbiomech.2006.12.010
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
Background. Distal radius fracture is among the most common type of skeletal injuries. To conquer the surgical and biomechanical complications of the most-frequent used double-plating operation for this fracture, modified double-plating technique was proposed in this study. The aim of this study was to investigate the biomechanical interactions of double-plating, modified double-plating and traditional single plating fixations coupled with various load conditions using nonlinear finite element analysis. Methods. A three-dimensional finite element distal radius fracture model with three fixation methods (double-plating, modified double-plating and single) was generated based on computer tomography data. After model verification and validation, frictional (contact) elements were used to simulate the interface condition between the fixation plates and the bony surface. The rigidity, stress values and displacements at the radius end were observed under axial, bending and torsion load conditions. Findings. The simulated results showed that the modified double-plating model demonstrated the highest rigidity and the least displacement among the three techniques in bending, but not in axial compression (similar results across the three) and torsion (modified double-plating technique possessed lowest rigidity). The maximum von Mises stress for bone was lower in modified double-plating model as well. These results indicated that modified double-plating technique demonstrated a better structural strength against bending with the least potential of fracture fragments and screw loosening. Interpretation. Although a lower torsional rigidity, modified double-plating technique was a better choice in distal radius fracture fixation since the bending force, which has the potential to separate the fracture ends, is more detrimental in hindering fracture healing. (C) 2007 Elsevier Ltd. All rights reserved.
引用
收藏
页码:510 / 517
页数:8
相关论文
共 25 条
[1]
Stress analysis of interbody fusion-finite element modelling of intervertebral implant and vertebral body [J].
Adam, C ;
Pearcy, M ;
McCombe, P .
CLINICAL BIOMECHANICS, 2003, 18 (04) :265-272
[2]
FORCES IN THE NORMAL AND ABNORMAL HAND [J].
AN, KN ;
CHAO, EY ;
COONEY, WP ;
LINSCHEID, RL .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1985, 3 (02) :202-211
[3]
ARO HT, 1993, CLIN ORTHOP RELAT R, P8
[4]
DEVELOPMENT OF A DEGRADABLE COMPOSITE FOR ORTHOPEDIC USE - MECHANICAL EVALUATION OF AN HYDROXYAPATITE POLYHYDROXYBUTYRATE COMPOSITE-MATERIAL [J].
BOEREE, NR ;
DOVE, J ;
COOPER, JJ ;
KNOWLES, JC ;
HASTINGS, GW .
BIOMATERIALS, 1993, 14 (10) :793-796
[5]
Intramedullary nails with two lag screws [J].
Brown, CJ ;
Wang, CJ ;
Yettram, AL ;
Procter, P .
CLINICAL BIOMECHANICS, 2004, 19 (05) :519-525
[6]
Displaced intra-articular fractures of the distal aspect of the radius - Long-term results in young adults after open reduction and internal fixation [J].
Catalano, LW ;
Cole, RJ ;
Gelberman, RH ;
Evanoff, BA ;
Gilula, LA ;
Borrelli, J .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (09) :1290-1302
[7]
RIGIDITY AND STRESS-ANALYSES OF EXTERNAL FRACTURE FIXATION DEVICES - A THEORETICAL APPROACH [J].
CHAO, EY ;
KASMAN, RA ;
AN, KN .
JOURNAL OF BIOMECHANICS, 1982, 15 (12) :971-&
[8]
Selection of fixation devices in proximal femur rotational osteotomy: clinical complications and finite element analysis [J].
Chen, WP ;
Tai, CL ;
Shih, CH ;
Hsieh, PH ;
Leou, MC ;
Lee, MS .
CLINICAL BIOMECHANICS, 2004, 19 (03) :255-262
[9]
Dunning CE, 1999, J HAND SURG-AM, V24A, P992
[10]
A biomechanical evaluation of lateral plating of distal radial shaft fractures [J].
Eglseder, WA ;
Jasper, LE ;
Davis, CW ;
Belkoff, SM .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2003, 28A (06) :959-963