摘要
The pattern electroretinogram (PERG) is a retinal response evoked by viewing a temporally alternating pattern, usually a black and white checkerboard or grating. The PERG is important in clinical and research applications because it provides information both about retinal ganglion cell function and, because the stimulus is customarily viewed with central fixation, the function of the macula. The PERG can therefore facilitate interpretation of an abnormal pattern VEP by revealing the retinal responses to a similar stimulus to that used for the VEP. However, practitioners may have difficulty choosing between the different techniques for recording the PERG that have been described in the literature. The International Society for Clinical Electrophysiology of Vision published a standard for clinical PERG recording in 2000 to assist practitioners in obtaining good quality reliable responses and to facilitate inter-laboratory communication and comparison. This document is the scheduled revision of that standard.
引用
收藏
页码:111 / 116
页数:6
相关论文
共 4 条
[1]   Standard for pattern electroretinography [J].
Bach M. ;
Hawlina M. ;
Holder G.E. ;
Marmor M.F. ;
Meigen T. ;
Vaegan ;
Miyake Y. .
Documenta Ophthalmologica, 2000, 101 (1) :11-18
[2]   Guidelines for calibration of stimulus and recording parameters used in clinical electrophysiology of vision [J].
Brigell, M ;
Bach, M ;
Barber, C ;
Kawasaki, K ;
Kooijman, A .
DOCUMENTA OPHTHALMOLOGICA, 1998, 95 (01) :1-14
[3]   Standard for clinical electroretinography (2004 update) [J].
Marmor, MF ;
Holder, GE ;
Seeliger, MW ;
Yamamoto, S .
DOCUMENTA OPHTHALMOLOGICA, 2004, 108 (02) :107-114
[4]  
MARMOR MF, 1996, DOC OPHTHALMOL, V91, P291