Thulium Fiber Laser Lithotripsy Using Tapered Fibers

被引:64
作者
Blackmon, Richard L. [1 ]
Irby, Pierce B. [2 ]
Fried, Nathaniel M. [1 ,3 ]
机构
[1] Univ N Carolina, Dept Phys & Opt Sci, Charlotte, NC 28223 USA
[2] Carolinas Med Ctr, Dept Urol, Charlotte, NC 28203 USA
[3] Johns Hopkins Med Inst, Dept Urol, Baltimore, MD 21205 USA
关键词
Thulium; lithotripsy; tapered fibers; urinary stones; HOLMIUM-YAG LITHOTRIPSY; OPTICAL-FIBERS;
D O I
10.1002/lsm.20883
中图分类号
R75 [皮肤病学与性病学];
学科分类号
100206 ;
摘要
Introduction: The Thulium fiber laser has recently been tested as a potential alternative to the Holmium:YAG laser for lithotripsy. This study explores use of a short taper for expanding the Thulium fiber laser beam at the distal tip of a small-core fiber. Methods: Thulium fiber laser radiation with a wavelength of 1,908 nm, 10 Hz pulse rate, 70 mJ pulse energy, and 1-millisecond pulse duration was delivered through a 2-m-length fiber with 150-mu m-core-input-end, 300-mu m-core-output-end, and 5-mm-length taper, in contact with human uric acid (UA) and calcium oxalate monohydrate (COM) stones, ex vivo (n = 10 each). Stone mass loss, stone crater depths, fiber transmission losses, fiber burn-back, irrigation rates, and deflection through a flexible ureteroscope were measured for the tapered fiber and compared with conventional fibers. Results: After delivery of 1,800 pulses through the tapered fiber, mass loss measured 12.7 +/- 2.6 mg for UA and 7.2 +/- 0.8 mg COM stones, comparable to conventional 100-mu m-core fibers (12.6 +/- 2.5 mg for UA and 6.8 +/- 1.7 mg for COM stones). No transmission losses or burn-back occurred for the tapered fiber after 36,000 pulses, while a conventional 150-mu m fiber experienced significant tip degradation after only 1,800 pulses. High irrigation rates were measured with the tapered fiber inserted through the working port of a flexible ureteroscope without hindering its deflection, mimicking that of a conventional 150 mu m fiber. Conclusions: The short tapered distal fiber tip allows expansion of the laser beam, resulting in decreased fiber tip damage compared to conventional small-core fibers, without compromising fiber bending, stone vaporization efficiency, or irrigation rates. Lasers Surg. Med. 42:45-50, 2010. (C) 2010 Wiley-Liss, Inc.
引用
收藏
页码:45 / 50
页数:6
相关论文
共 11 条
[1]   Comparison of flexible ureteroscopes: Deflection, irrigant flow and optical characteristics [J].
Abdelshehid, C ;
Ahlering, MT ;
Chou, D ;
Park, HK ;
Basillote, J ;
Lee, D ;
Kim, I ;
Eichel, L ;
Protsenko, D ;
Wong, B ;
McDougall, E ;
Clayman, RV .
JOURNAL OF UROLOGY, 2005, 173 (06) :2017-2021
[2]  
BLACKMON RL, P SPIE IN PRESS, P14031
[3]   Shaped fiber tips for medical and industrial applications [J].
Clarkin, JP ;
Timmerman, RJ ;
Shannon, JH .
OPTICAL FIBERS AND SENSORS FOR MEDICAL APPLICATIONS IV, 2004, 5317 :70-80
[4]   Thulium fiber laser lithotripsy:: An in vitro analysis of stone fragmentation using a modulated 110-watt thulium fiber laser at 1.94 μm [J].
Fried, NM .
LASERS IN SURGERY AND MEDICINE, 2005, 37 (01) :53-58
[5]   Performance and safety of holmium: YAG laser optical fibers [J].
Knudsen, BE ;
Glickman, RD ;
Stallman, KJ ;
Maswadi, S ;
Chew, BH ;
Beiko, DT ;
Denstedt, JD ;
Teichman, JMH .
JOURNAL OF ENDOUROLOGY, 2005, 19 (09) :1092-1097
[6]   Quantification of Holmium: Yttrium Aluminum Garnet Optical Tip Degradation [J].
Mues, Adam C. ;
Teichman, Joel M. H. ;
Knudsen, Bodo E. .
JOURNAL OF ENDOUROLOGY, 2009, 23 (09) :1425-1428
[7]   Evaluation of 24 Holmium:YAG Laser Optical Fibers for Flexible Ureteroscopy [J].
Mues, Adam C. ;
Teichman, Joel M. H. ;
Knudsen, Bodo E. .
JOURNAL OF UROLOGY, 2009, 182 (01) :348-354
[8]   Review of laser fibers: A practical guide for urologists [J].
Nazif, OA ;
Teichman, JMH ;
Glickman, RD ;
Welch, AJ .
JOURNAL OF ENDOUROLOGY, 2004, 18 (09) :818-829
[9]   Thulium Fiber Laser Ablation of Urinary Stones Through Small-Core Optical Fibers [J].
Scott, Nicholas J. ;
Cilip, Christopher M. ;
Fried, Nathaniel M. .
IEEE JOURNAL OF SELECTED TOPICS IN QUANTUM ELECTRONICS, 2009, 15 (02) :435-440
[10]   Holmium:YAG lithotripsy:: Optimal power settings [J].
Spore, SS ;
Teichman, JMH ;
Corbin, NS ;
Champion, PC ;
Williamson, EA ;
Glickman, RD .
JOURNAL OF ENDOUROLOGY, 1999, 13 (08) :559-566