Influence of PEEK Coating on Hip Implant Stress Shielding: A Finite Element Analysis

被引:51
作者
Anguiano-Sanchez, Jesica [1 ]
Martinez-Romero, Oscar [1 ]
Siller, Hector R. [1 ]
Diaz-Elizondo, Jose A. [2 ]
Flores-Villalba, Eduardo [1 ,2 ]
Rodriguez, Ciro A. [1 ]
机构
[1] Tecnol Monterrey, Escuela Ingn & Ciencias, Ave Eugenio Garza Sada 2501, Monterrey 64849, Mexico
[2] Tecnol Monterrey, Escuela Med, Ave Eugenio Garza Sada 2501, Monterrey 64849, Mexico
关键词
ELECTROPHORETIC DEPOSITION; DESIGN; PROSTHESES; FRACTURE; STEM;
D O I
10.1155/2016/6183679
中图分类号
Q [生物科学];
学科分类号
090105 [作物生产系统与生态工程];
摘要
Stress shielding is a well-known failure factor in hip implants. This work proposes a design concept for hip implants, using a combination of metallic stem with a polymer coating (polyether ether ketone (PEEK)). The proposed design concept is simulated using titanium alloy stems and PEEK coatings with thicknesses varying from 100 to 400 mu m. The Finite Element analysis of the cancellous bone surrounding the implant shows promising results. The effective von Mises stress increases between 81 and 92% for the complete volume of cancellous bone. When focusing on the proximal zone of the implant, the increased stress transmission to the cancellous bone reaches between 47 and 60%. This increment in load transferred to the bone can influence mineral bone loss due to stress shielding, minimizing such effect, and thus prolonging implant lifespan.
引用
收藏
页数:10
相关论文
共 24 条
[1]
An experimental study of process variables in turning operations of Ti-6Al-4V and Cr-Co spherical prostheses [J].
Abellan-Nebot, J. V. ;
Siller, H. R. ;
Vila, C. ;
Rodriguez, C. A. .
INTERNATIONAL JOURNAL OF ADVANCED MANUFACTURING TECHNOLOGY, 2012, 63 (9-12) :887-902
[2]
Investigating stress shielding spanned by biomimetic polymer-composite vs. metallic hip stem: A computational study using mechano-biochemical model [J].
Avval, Pouria Tavakkoli ;
Samiezadeh, Saeid ;
Klika, Vaclav ;
Bougherara, Habiba .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2015, 41 :56-67
[3]
A review on fundamentals and applications of electrophoretic deposition (EPD) [J].
Besra, Laxmidhar ;
Liu, Meilin .
PROGRESS IN MATERIALS SCIENCE, 2007, 52 (01) :1-61
[4]
Design of a biomimetic polymer-composite hip prosthesis [J].
Bougherara, Habiba ;
Bureau, Martin ;
Campbell, Melissa ;
Vadean, Aurelian ;
Yahia, L'Hocine .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2007, 82A (01) :27-40
[5]
Comparison of different hip prosthesis shapes considering micro-level bone remodeling and stress-shielding criteria using three-dimensional design space topology optimization [J].
Boyle, Christopher ;
Kim, Il Yong .
JOURNAL OF BIOMECHANICS, 2011, 44 (09) :1722-1728
[6]
Design and analysis of robust total joint replacements: Finite element model experiments with environmental variables [J].
Chang, PB ;
Williams, BJ ;
Bhalla, KSB ;
Belknap, TW ;
Santner, TJ ;
Notz, WI ;
Bartel, DL .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2001, 123 (03) :239-246
[7]
Electrophoretic deposition of PEEK-nano alumina composite coatings on stainless steel [J].
Corni, Ilaria ;
Neumann, Nico ;
Novak, Sasa ;
Koenig, Katja ;
Veronesi, Paolo ;
Chen, Qizhi ;
Ryan, Mary P. ;
Boccaccini, Aldo R. .
SURFACE & COATINGS TECHNOLOGY, 2009, 203 (10-11) :1349-1359
[8]
Femoral neck fracture following hip resurfacing THE EFFECT OF ALIGNMENT OF THE FEMORAL COMPONENT [J].
Davis, E. T. ;
Olsen, M. ;
Zdero, R. ;
Waddell, J. P. ;
Schemitsch, E. H. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2008, 90B (11) :1522-1527
[9]
Modelling bone tissue fracture and healing:: a review [J].
Doblaré, M ;
García, JM ;
Gómez, MJ .
ENGINEERING FRACTURE MECHANICS, 2004, 71 (13-14) :1809-1840
[10]
Biomechanical properties of an intact, injured, repaired, and healed femur: An experimental and computational study [J].
Ebrahimi, Hamid ;
Rabinovich, Mikhail ;
Vuleta, Viktor ;
Zalcman, Daniel ;
Shah, Suraj ;
Dubov, Anton ;
Roy, Kavya ;
Siddiqui, Faisal Sharaf ;
Schemitsch, Emil H. ;
Bougherara, Habiba ;
Zdero, Rad .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2012, 16 :121-135