Immunoaffinity chromatography combined on-line with high-performance liquid chromatography mass spectrometry for the determination of corticosteroids

被引:41
作者
Creaser, CS
Feely, SJ
Houghton, E
Seymour, M
机构
[1] Nottingham Trent Univ, Dept Chem & Phys, Nottingham NG11 8NS, England
[2] Horseracing Forens Lab Ltd, Fordham CB7 5WP, Cambs, England
基金
英国工程与自然科学研究理事会;
关键词
immunoaffinity chromatography; corticosteroids; dexamethasone; flumethasone;
D O I
10.1016/S0021-9673(97)00888-1
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
On-line coupled immunoaffinity chromatography-reversed-phase high-performance liquid chromatography (IAC-HPLC) with detection by quadrupole ion trap mass spectrometry using a particle beam interface has been developed for the determination of the steroids, dexamethasone and flumethasone. HEMA (polyhydroxyethylmethacrylate) was evaluated as a support material for the anti-dexamethasone antibodies used in IAC. Antibody cross-reactivity and non-specific binding have been investigated for the HEMA bound anti-dexamethasone IAC column. The on-line IAC-HPLC-MS determination of dexamethasone and flumethasone in post-administration equine urine samples showed precisions (R.S.D.) of 8.0 and 7.1%, respectively, with limits of detection in the range 3-4 ng/ml. (C) 1998 Elsevier Science B.V.
引用
收藏
页码:37 / 43
页数:7
相关论文
共 26 条