Prospective evaluation of automated follicle monitoring in 58 in vitro fertilization cycles: follicular volume as a new indicator of oocyte maturity

被引:26
作者
Rodrigue-Fuentes, Adela [1 ]
Hernandez, Jairo [1 ]
Garcia-Guzman, Rocio [1 ]
Chinea, Elena [1 ]
Iaconianni, Laura [2 ]
Palumbo, Angela [1 ]
机构
[1] Ctr Asistencia Reprod Humana Canarias, E-38240 Tenerife, Spain
[2] Ctr Ecograf EcoBI, Rome, Italy
关键词
Follicular monitoring; three-dimensional ultrasound; SonoAVC; follicular volume; in vitro fertilization; SONOGRAPHY; ULTRASOUND;
D O I
10.1016/j.fertnstert.2009.02.058
中图分类号
R71 [妇产科学];
学科分类号
100211 [妇产科学];
摘要
Objectives: To assess the practical use of SonoAVC in an IVF program, and to establish new criteria for hCG administration based on follicular volume. Design: Prospective clinical study. Setting: Private IVF Center. Patient(s): Fifty-eight women with infertility undergoing IVF. Intervention(s): Two dimensional (2D) and three dimensional (3D) scanning on the day of hCG administration. Main Outcome Measure(s): image quality, mean follicular diameter obtained by 2D and 3D sonography, follicular volume, number of oocytes retrieved, number of mature oocytes, time needed for each examination. Result(s): Approximately 60% of the patients included in this study had good image quality and could be monitored by 3D scans with subsequent application of the SonoAVC software. When image quality is good, measurements obtained by the automated mode are comparable to those obtained manually in 62% of cases. Automated monitoring is significantly quicker than conventional manual monitoring. Follicles with a measured volume >= 0.6 cc on the day of hCG administration are associated with the finding of mature oocytes at the time of egg retrieval. Conclusion(s): SonoAVC allows reliable evaluation of stimulated ovaries, and may help us establish new criteria for timing hCG administration based on follicular volume estimation rather than follicular size. Software improvements are needed to improve universal patient use. (Fertil Steril(R) 2010-793:616-20. (C)2010 by American Society for Reproductive Medicine.)
引用
收藏
页码:616 / 620
页数:5
相关论文
共 7 条
[1]
Deutch TD, 2007, FERTIL STERIL, V88, pS80
[2]
Automated follicle tracking facilitates standardization and may improve work flow [J].
Raine-Fenning, N. ;
Jayaprakasan, K. ;
Clewes, J. .
ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 2007, 30 (07) :1015-1018
[3]
The interobserver reliability and validity of volume calculation from three-dimensional ultrasound datasets in the in vitro setting [J].
Raine-Fenning, NJ ;
Clewes, JS ;
Kendall, NR ;
Bunkheila, AK ;
Campbell, BK ;
Johnson, IR .
ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 2003, 21 (03) :283-291
[4]
Salama S, 2008, FERTIL STERIL, V90, pS20, DOI [10.1016/j.fertnstert.2008.07.400, DOI 10.1016/J.FERTNSTERT.2008.07.400]
[5]
SHAPIRO BS, 1989, J REPROD MED, V34, P151
[6]
THE TECHNIQUE OF TRANS-VAGINAL SONOGRAPHY WITH THE USE OF A 6.5 MHZ PROBE [J].
TIMORTRITSCH, IE ;
BARYAM, Y ;
ELGALI, S ;
ROTTEM, S .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1988, 158 (05) :1019-1024
[7]
WITTMAACK FM, 1994, FERTIL STERIL, V62, P1205