The snuffbox technique: A reliable color Doppler method to assess hand circulation

被引:12
作者
Kochi, K [1 ]
Orihashi, K [1 ]
Sueda, T [1 ]
机构
[1] Hiroshima Univ, Sch Med, Dept Surg 1, Hiroshima, Japan
关键词
D O I
10.1067/mtc.2003.325
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: Although the Allen test is crude and subjective, an objective color Doppler method has yet to be established in the assessment of hand circulation before radial artery harvesting. Doppler studies so far have neglected the Doppler principle that the insonation angles should be less than 30degrees and have not been compared with any standard except the crude Allen test. We therefore introduced the snuffbox technique, measuring the radial artery at the anatomic snuffbox, which is the most distal area after harvesting. Color Doppler methods were compared with the stump pressure as a criterion standard. Methods: Maximal flow velocity and Doppler angles of the snuffbox, palmar artery, and ulnar artery were measured in 20 hands before and after radial artery harvesting. Stump pressure was measured during surgery. Results: Maximal flow velocity in the snuffbox decreased with radial artery compression and after radial artery harvesting. All the flow patterns in the snuffbox technique showed simultaneous reversal. Maximal flow velocity in the ulnar artery increased with radial artery compression and after radial artery harvesting. Maximal flow velocity in the palmar artery did not change significantly with radial artery compression or after radial artery harvesting. Doppler angles were 20.9degrees +/- 6.0degrees in the radial artery of the snuffbox, 82.5degrees +/- 6.1degrees at the palmar artery, and 81.0degrees +/- 7.6degrees at the ulnar artery. The changes in the snuffbox technique were highly correlated with the stump pressure ratio (P < .001). Conclusions: Among various color Doppler methods, the snuffbox technique was precise and reliable.
引用
收藏
页码:821 / 825
页数:5
相关论文
共 18 条
[1]  
ACAR C, 1999, ISCHAEMIC HEART DIS, P151
[2]  
AGUR AMR, 1991, GRANTS ATLAS ANATOMY, P353
[3]   The Allen test [J].
Cable, DG ;
Mullany, CJ ;
Schaff, HV .
ANNALS OF THORACIC SURGERY, 1999, 67 (03) :876-877
[4]   RETRACTED: Acute upper limb ischemia: A complication of coronary artery bypass grafting (Retracted Article) [J].
Fox, AD ;
Whiteley, MS ;
Phillips-Hughes, J ;
Roake, J .
ANNALS OF THORACIC SURGERY, 1999, 67 (02) :535-536
[5]   ARTERIAL PATTERNS IN THE HAND BASED ON A 3-DIMENSIONAL ANALYSIS OF 220 CADAVER HANDS [J].
IKEDA, A ;
UGAWA, A ;
KAZIHARA, Y ;
HAMADA, N .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 1988, 13A (04) :501-509
[6]   Reliability of Allen's test in selection of patients for radial artery harvest [J].
Jarvis, MA ;
Jarvis, CL ;
Jones, PRM ;
Spyt, TJ .
ANNALS OF THORACIC SURGERY, 2000, 70 (04) :1362-1365
[7]  
KAMIENSKI RW, 1976, SURG GYNECOL OBSTET, V142, P861
[8]   New noninvasive test alternative to Allen's test: Snuff-box technique [J].
Kochi, K ;
Sueda, T ;
Orihashi, K ;
Matsuura, Y .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1999, 118 (04) :756-758
[9]   An unexpected complication after harvesting of the radial artery for coronary artery bypass grafting [J].
Nunoo-Mensah, J .
ANNALS OF THORACIC SURGERY, 1998, 66 (03) :929-931
[10]   Safe removal of the radial artery for myocardial revascularization: A Doppler study to prevent ischemic complications to the hand [J].
Pola, P ;
Serricchio, M ;
Flore, R ;
Manasse, E ;
Favuzzi, A ;
Possati, GF .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1996, 112 (03) :737-744