Full-foil x-ray mapping of integrated column density applied to the absolute determination of mass attenuation coefficients

被引:21
作者
de Jonge, MD [1 ]
Barnea, Z [1 ]
Tran, CQ [1 ]
Chantler, CT [1 ]
机构
[1] Univ Melbourne, Sch Phys, Parkville, Vic 3052, Australia
关键词
integrated column density; full-foil mapping; x-ray; mass attenuation coefficient;
D O I
10.1088/0957-0233/15/9/019
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Recent measurements of mass attenuation coefficients have identified the determination of the thickness of the absorbing specimen as the major limitation to the accuracy of the measurement. We present a technique for determining the mass attenuation coefficient with high accuracy. The technique uses the integral of the density along a column extending through the thickness of the absorber, which we term the integrated column density. Attenuation measurements mapped across the entire absorber are used to determine a relative map of the integrated column density. These relative measurements are then placed on an absolute scale by comparison with the average integrated column density and are used to determine the mass attenuation coefficient. This approach correctly treats variations in the integrated column density across the foil. We illustrate the technique with an absolute measurement of the x-ray mass attenuation coefficient of molybdenum using a synchrotron beam of energy 41.568 keV +/- 0.005 keV. We obtain [mu/rho] = 11.6514 cm(2) g(-1) +/- 0.0032 cm(2) g(-1), accurate to 0.028%-over one order of magnitude more accurate than any previous work. The full-foil technique used to determine the mass attenuation coefficient is used to determine an integrated column density profile of a sample to a precision of around 0.05% of the thickness of the absorber. We demonstrate the sensitivity of the technique by observing a periodic thickness variation of order 0.1 mum occurring over a 5 mm length scale on a nominally 50 mum thick molybdenum foil.
引用
收藏
页码:1811 / 1822
页数:12
相关论文
共 46 条
[1]   Measurement of the mass attenuation coefficient for elemental materials in the range 6 ≤, Z ≤ 82 using X-rays from 13 up to 50 kev [J].
Angelone, M ;
Bubba, T ;
Esposito, A .
APPLIED RADIATION AND ISOTOPES, 2001, 55 (04) :505-511
[2]   Determination of X-ray photoelectric absorption of Ge and Si avoiding solid-state effects [J].
Baltazar-Rodrigues, J ;
Cusatis, C .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2001, 179 (03) :325-333
[3]   MULTIPLE BRAGG REFLECTION MONOCHROMATORS FOR SYNCHROTRON-X RADIATION [J].
BEAUMONT, JH ;
HART, M .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1974, 7 (10) :823-829
[4]   X-RAY ATTENUATION COEFFICIENTS OF GRAPHITE IN THE RANGE O-40-A TO 1-54-A [J].
BERRY, AA ;
LAWRENCE, JL .
ACTA CRYSTALLOGRAPHICA SECTION A, 1979, 35 (MAR) :316-318
[5]   CORE EXCITONS AND INNER WELL RESONANCES IN SURFACE SOFT-X-RAY ABSORPTION (SSXA) SPECTRA [J].
BIANCONI, A .
SURFACE SCIENCE, 1979, 89 (1-3) :41-50
[6]   PERFECT-CRYSTAL MONOCHROMATORS FOR SYNCHROTRON X-RADIATION [J].
BONSE, U ;
MATERLIK, G ;
SCHRODER, W .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1976, 9 (JUN1) :223-230
[7]   DENSITY SCALE BASED ON SOLID OBJECTS [J].
BOWMAN, HA ;
SCHOONOVER, RM ;
CARROLL, CL .
JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS SECTION A-PHYSICS AND CHEMISTRY, 1974, A 78 (01) :13-40
[8]   The effect of metal oxide on the no-flow underfill fluxing capacity [J].
Chan, YC ;
Fan, SH ;
Hung, KC .
MATERIALS SCIENCE AND ENGINEERING B-SOLID STATE MATERIALS FOR ADVANCED TECHNOLOGY, 2001, 85 (01) :64-69
[9]  
Chantler CT, 2000, X-RAY SPECTROM, V29, P459, DOI 10.1002/1097-4539(200011/12)29:6<459::AID-XRS453>3.0.CO
[10]  
2-N