Single crystal PZN/PT-polymer composites for ultrasound transducer applications

被引:147
作者
Ritter, T
Geng, XC
Shung, KK
Lopath, PD
Park, SE
Shrout, TR
机构
[1] Penn State Univ, NIH Resource Ctr Med Ultrason Transducer Technol, University Pk, PA 16802 USA
[2] Blatek Inc, State Coll, PA USA
关键词
D O I
10.1109/58.852060
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Single crystal relaxer ferroelectrics of PZN-8%PT were investigated for potential application in ultrasound transducers. The full set of electromechanical properties was determined using combined resonance and laser interferometry techniques. Ultra-high length extensional coupling (k(33)) of 0.94 was observed, a 25% increase over Navy Type VI PZT ceramics. The thickness extensional coupling (k(t)) of 0.48 was comparable to PZT compositions, and the compliance S-33(E) was a factor of six greater. To maximize height extensional coupling (k'(33)), while minimizing length extensional coupling k(31) in array elements, it was necessary to align the elements along the [100] crystallographic direction in the x-y plane. Mode coupling plots and test samples for array elements determined that width-to-height ratios of less than 0.5 were desired, similar to the requirement for polycrystalline PZT ceramics. Modeling of 1-3 composites and experimental results demonstrated that thickness coupling greater than 0.80 could be achieved with a 40% to 70% volume fraction of PZN-PT, Although this is a substantial increase over PZT 1-3 composites, with a thickness coupling coefficient of 0.66, it represents a smaller fraction of the length extensional coupling k33 This reduction may be a consequence of the increased compliance of PZN-PT, which results in significant clamping by the polymer matrix. Ultrasonic transducers fabricated using PZN-8%PT 1-3 composites achieved experimental bandwidths as high as 141%, The pulse-echo responses displayed good agreement with modeled results using the Redwood equivalent circuit.
引用
收藏
页码:792 / 800
页数:9
相关论文
共 26 条
[1]  
[Anonymous], 1987, ACOUSTIC WAVES DEVIC
[2]  
Auld B.A., 1973, ACOUSTIC FIELDS WAVE
[3]   DESIGN OF EFFICIENT BROAD-BAND PIEZOELECTRIC TRANSDUCERS [J].
DESILETS, CS ;
FRASER, JD ;
KINO, GS .
IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1978, 25 (03) :115-125
[4]  
DESILETS CS, 1978, TRANSDUCER ARRAYS SU
[5]  
GENG X, 1999, J APPL PHYS, V85, P1
[6]   Optimization of 1-3 relaxor-PT single crystal polymer composite [J].
Geng, XC ;
Ritter, T ;
Shung, KK .
MEDICAL IMAGING 1999: ULTRASONIC TRANSDUCER ENGINEERING, 1999, 3664 :24-30
[7]  
Geng XC, 1998, ULTRASON, P571, DOI 10.1109/ULTSYM.1998.762215
[8]   PIEZOELECTRIC COMPOSITE-MATERIALS FOR ULTRASONIC TRANSDUCER APPLICATIONS .1. RESONANT MODES OF VIBRATION OF PZT ROD POLYMER COMPOSITES [J].
GURURAJA, TR ;
SCHULZE, WA ;
CROSS, LE ;
NEWNHAM, RE ;
AULD, BA ;
WANG, YJ .
IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1985, 32 (04) :481-498
[9]  
Haun M. J., 1986, Ferroelectrics, V68, P123, DOI 10.1080/00150198608238743
[10]   A THEORETICAL-STUDY ON THE INFLUENCE OF SOME CONSTITUENT MATERIAL PROPERTIES ON THE BEHAVIOR OF 1-3 CONNECTIVITY COMPOSITE TRANSDUCERS [J].
HAYWARD, G ;
BENNETT, J ;
HAMILTON, R .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1995, 98 (04) :2187-2196