Performance evaluation of a multi-slice CT system with 16-slice detector and increased gantry rotation speed for isotropic submillimeter imaging of the heart

被引:17
作者
Flohr, T
Küttner, A
Bruder, H
Stierstorfer, K
Halliburton, SS
Schaller, S
Ohnesorge, BM
机构
[1] Siemens Med Solut, CTC CF, D-91301 Forchheim, Germany
[2] Univ Tubingen, Dept Diagnost Radiol, Tubingen, Germany
[3] Cleveland Clin Fdn, Cleveland, OH 44195 USA
关键词
cardiac imaging; multi-slice CT; ECG-gating; coronary CTA; NONINVASIVE CORONARY-ANGIOGRAPHY; COMPUTED-TOMOGRAPHY; SPIRAL CT; TECHNICAL DEVELOPMENTS; TEMPORAL RESOLUTION; ELECTRON-BEAM; HELICAL CT; RECONSTRUCTION; OPTIMIZATION; PLAQUES;
D O I
10.1007/s00059-003-2456-1
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: 4-slice CT scanners have shown limitations in clinical application for noninvasive coronary CT angiography (CTA). We evaluate advances in ECG-gated scanning of the heart and the coronary arteries with recently introduced 16-slice CT equipment (SOMATOM Sensation 16, Siemens, Forchheim, Germany). Materials and Methods: The technical principles of ECG-gated cardiac scanning, scan parameters, and detector design of the new scanner are presented. ECG-gated scan and image reconstruction techniques and ECG-controlled dose modulation ("ECG pulsing") for a reduction of the patient dose are described, key parameters for image quality and simulation results presented, and phantom studies and initial patient experience discussed. The impact of reduced gantry rotation time (0-42 s) on temporal resolution and initial estimations of the patient dose are presented. Results: Extensions of ECG-gated reconstruction algorithms used for 4-slice CT provide adequate image quality for up to 16 slices. For each detector collimation different slice widths are available for retrospective reconstruction with well-defined slice sensitivity profiles (SSPs). For coronary CTA the heart can be covered with 0.75 mm collimation within a 20-s breathhold. The best possible spatial resolution is 0.5 X 0.5 X 0.6 mm. For 0.42 s gantry rotation time, temporal resolution reaches its Optimum (105 ms) at a heart rate of 81 bpm. Effective patient dose for coronary CTA is 4-5 mSv using ECG-pulsed acquisition. Conclusion: The clinical performance of coronary CTA by means of spatial resolution, temporal resolution and scan time is substantially improved with the evaluated 16-slice CT scanner. Also, display of smaller coronary segments and instent visualization are substantially improved.
引用
收藏
页码:7 / +
页数:13
相关论文
共 28 条
[1]  
Achenbach S, 2000, CIRCULATION, V102, P2823
[2]   Curved multiplanar reconstructions for the evaluation of contrast-enhanced electron beam CT of the coronary arteries [J].
Achenbach, S ;
Moshage, W ;
Ropers, D ;
Bachmann, K .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1998, 170 (04) :895-899
[3]  
BAHNER ML, 1999, EUR RADIOL, V9, P106
[4]   Imaging of noncalcified coronary plaques using helical CT with retrospective ECG gating [J].
Becker, CR ;
Knez, A ;
Ohnesorge, B ;
Schoepf, UJ ;
Reiser, MF .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 175 (02) :423-424
[5]   Detection and quantification of coronary artery calcification with electron-beam and conventional CT [J].
Becker, CR ;
Knez, A ;
Jakobs, TF ;
Aydemir, S ;
Becker, A ;
Schoepf, UJ ;
Bruening, R ;
Haberl, R ;
Reiser, MF .
EUROPEAN RADIOLOGY, 1999, 9 (04) :620-624
[6]   Ultrafast computed tomography as a diagnostic modality in the detection of coronary artery disease: A multicenter study [J].
Budoff, MJ ;
Georgiou, D ;
Brody, A ;
Agatston, AS ;
Kennedy, J ;
Wolfkiel, C ;
Stanford, W ;
Shields, P ;
Lewis, RJ ;
Janowitz, WR ;
Rich, S ;
Brundage, BH .
CIRCULATION, 1996, 93 (05) :898-904
[7]   New technical developments in multislice CT - Part 1: Approaching isotropic resolution with sub-millimeter 16-slice scanning [J].
Flohr, T ;
Stierstorfer, K ;
Bruder, H ;
Simon, J ;
Schaller, S .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2002, 174 (07) :839-845
[8]   New technical developments in multislice CT, part 2: Sub-millimeter 16-slice scanning and increased gantry rotation speed for cardiac Imaging [J].
Flohr, T ;
Bruder, H ;
Stierstorfer, K ;
Simon, J ;
Schaller, S ;
Ohnesorge, B .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2002, 174 (08) :1022-1027
[9]   Heart rate adaptive optimization of spatial and temporal resolution for electrocardiogram-gated multislice spiral CT of the heart [J].
Flohr, T ;
Ohnesorge, B .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2001, 25 (06) :907-923
[10]  
FLOHR T, 2000, 86 SCI ASS ANN M R P, V217, P438