Generation of synthetic background spectra by filtering the sample interferogram in FT-IR

被引:8
作者
Espinoza, LH
Niemczyk, TM [1 ]
Stallard, BR
机构
[1] Univ New Mexico, Dept Chem, Albuquerque, NM 87131 USA
[2] Sandia Natl Labs, Albuquerque, NM 87185 USA
关键词
fourier transform infrared spectroscopy; gasphase infrared spectroscopy; background correction; synthetic background;
D O I
10.1366/0003702981943554
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The calculation of an absorbance spectrum depends on the measurement of a blank, or background spectrum. In many cases, such as the determination of atmospheric constituents with the use of open-path Fourier transform infrared spectroscopy (FT-IR) or the determination of water vapor in a gaseous sample, it is very difficult to obtain a good background spectrum. The difficulty is due to the fact that it is nearly impossible in these situations to measure a spectrum with no analyte features present. We present a method of generating a background spectrum based on filtering the analyte features from the sample spectrum. When the filtering method is used, the accuracy of the results obtained is found to be dependent upon the analyte peak width, peak height, and type of filter employed. Guidelines for the use of this background generation technique for quantitative determinations are presented.
引用
收藏
页码:375 / 379
页数:5
相关论文
共 8 条
[1]  
CARTER RE, 1993, J AIR WASTE MANAGE, V43, P480
[2]  
FATELEY WG, 1992, SPIE P, V1637, P50
[3]   IMPROVED SENSITIVITY OF INFRARED-SPECTROSCOPY BY THE APPLICATION OF LEAST-SQUARES METHODS [J].
HAALAND, DM ;
EASTERLING, RG .
APPLIED SPECTROSCOPY, 1980, 34 (05) :539-548
[4]  
Hanst P. L., 1993, SPECTROSCOPY, V8, P44
[5]   DETECTION OF ATMOSPHERIC POLLUTANTS AT PARTS-PER-BILLION LEVELS BY INFRARED SPECTROSCOPY [J].
HANST, PL ;
LEFOHN, AS ;
GAY, BW .
APPLIED SPECTROSCOPY, 1973, 27 (03) :188-198
[6]   DIGITIZATION NOISE IN FOURIER-TRANSFORM INFRARED SPECTROSCOPY AND GAIN RANGING [J].
HIRSCHFELD, T .
APPLIED SPECTROSCOPY, 1979, 33 (05) :525-527
[7]  
RUSSWURM GM, 1995, LASER FOCUS WORLD, V31, P79
[8]   TRACE WATER-VAPOR DETECTION IN NITROGEN AND CORROSIVE-GASES BY FTIR SPECTROSCOPY [J].
STALLARD, BR ;
ESPINOZA, LH ;
ROWE, RK ;
GARCIA, MJ ;
NIEMCZYK, TM .