A comparison of simulator-tested and -retrieved cervical disc prostheses

被引:44
作者
Anderson, PA
Rouleau, JP
Toth, JM
Riew, KD
机构
[1] Univ Wisconsin, Dept Orthoped Surg & Rehabil, Madison, WI 53792 USA
[2] Med Coll Wisconsin, Dept Orthopaed Surg, Milwaukee, WI 53226 USA
[3] Medtronic Inc, Dept Sci & Technol, Minneapolis, MN USA
[4] Barnes Jewish Hosp, Dept Orthopaed Surg, St Louis, MO 63110 USA
[5] Washington Univ, Sch Med, St Louis, MO 63110 USA
关键词
cervical spine; disc degeneration; explant analysis; artificial disc replacement; disc arthroplasty;
D O I
10.3171/spi.2004.1.2.0202
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Object. Total joint arthroplasties most commonly fail because the implant becomes worn and a host inflammatory response subsequently develops. Both the material response to the biological environment and the host response to the device must be thoroughly evaluated to establish the efficacy of cervical arthroplasty. Analyses of devices explanted in humans allow evaluation of both responses. Hypothetical wear rates can be determined by comparing in vivo wear with simulator-derived wear. The purpose of this study was to perform explant analyses involving the Bryan and Prestige discs and compare these results with those obtained using spine simulators. Methods. Of the approximately 5500 patients treated with the Bryan disc, 11 have undergone explantation of the device. Six of these devices were analyzed for dimensional and chemical changes. Three of the approximately 300 implanted Prestige discs were retrieved, and two were examined microscopically. Histological specimens were assessed for wear particles and host inflammatory response. Additionally, the extent of simulator-produced wear was compared with that demonstrated in the retrieved specimens. Conclusions. The simulator-generated results predict adequate wear-related characteristics for both the Bryan and Prestige prostheses for a minimum of 40 years. Comparison of data with those of the retrieved specimens indicates that the wear was more minimal than predicted in simulators by five- to 10-fold. In no instance did the revisions result from failure of the device due to a reaction to wear debris, fracture, polymer oxidation, or metal corrosion. The inflammatory response seen in the periprosthetic tissues was minimal and not characteristic of inflammatory responses in failed diarthrodial joint arthroplasties.
引用
收藏
页码:202 / 210
页数:9
相关论文
共 19 条
[1]
Wear analysis of the Bryan Cervical Disc prosthesis [J].
Anderson, PA ;
Rouleau, JP ;
Bryan, VE ;
Carlson, CS .
SPINE, 2003, 28 (20) :S186-S194
[2]
[Anonymous], 1999, ANN BOOK ASTM STAND, V03.02
[3]
Active range of motion utilized in the cervical spine to perform daily functional tasks [J].
Bennett, SE ;
Schenk, RJ ;
Simmons, ED .
JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2002, 15 (04) :307-311
[4]
Cervical motion segment replacement [J].
Bryan, VE .
EUROPEAN SPINE JOURNAL, 2002, 11 (Suppl 2) :S92-S97
[5]
Positive cytokine production in failed metal-on-metal total hip replacements [J].
Campbell, PA ;
Wang, M ;
Amstutz, HC ;
Goodman, SB .
ACTA ORTHOPAEDICA SCANDINAVICA, 2002, 73 (05) :506-512
[6]
Surgical experience with an implanted artificial cervical joint [J].
Cummins, BH ;
Robertson, JT ;
Gill, SS .
JOURNAL OF NEUROSURGERY, 1998, 88 (06) :943-948
[7]
AGE AND GENDER RELATED NORMAL MOTION OF THE CERVICAL-SPINE [J].
DVORAK, J ;
ANTINNES, JA ;
PANJABI, M ;
LOUSTALOT, D ;
BONOMO, M .
SPINE, 1992, 17 (10) :S393-S398
[8]
Intermediate follow-up after treatment of degenerative disc disease with the Bryan Cervical Disc Prosthesis: Single-level and bi-level [J].
Goffin, J ;
Van Calenbergh, F ;
van Loon, J ;
Casey, A ;
Kehr, P ;
Liebig, K ;
Lind, B ;
Logroscino, C ;
Sgrambiglia, R ;
Pointillart, V .
SPINE, 2003, 28 (24) :2673-2678
[9]
A multi-station hip joint simulator study of the performance of 22 mm diameter zirconia-ultra-high molecular weight polyethylene total replacement hip joints [J].
Goldsmith, AAJ ;
Dowson, D .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1999, 213 (H2) :77-90
[10]
Cellular profile and cytokine production at prosthetic interfaces - Study of tissues retrieved from revised hip and knee replacements [J].
Goodman, SB ;
Huie, P ;
Song, Y ;
Schurman, D ;
Maloney, W ;
Woolson, S ;
Sibley, R .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1998, 80B (03) :531-539