Thyroid volume measurement in thyrotoxic patients: comparison between ultrasonography and iodine-124 positron emission tomography

被引:40
作者
Crawford, DC
Flower, MA
Pratt, BE
Hill, C
Zweit, J
McCready, R
Harmer, CL
机构
[1] Royal Marsden NHS Trust, Sutton, Surrey, England
[2] Canc Res Inst, Joint Dept Phys, Sutton, Surrey, England
[3] Royal Marsden NHS Trust, Dept Radiol, Sutton, Surrey, England
[4] Royal Marsden NHS Trust, Dept Nucl Med, Sutton, Surrey, England
[5] Royal Marsden NHS Trust, Dept Radiotherapy, Sutton, Surrey, England
关键词
thyroid volume; hyperthyroidism; ultrasonography; iodine-124 positron emission tomography; iodine-131; therapy;
D O I
10.1007/s002590050176
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this paper was to compare ultrasound (US) assessment of thyroid volume with that obtained by positron emission tomography (PET), in patients scheduled for adaptive radioiodine therapy, in which 50 Gy was prescribed to the functional PET volume, Firstly a pilot study was performed to ascertain the optimum method for US assessment of thyroid volume, Then 17 comparative measurements of thyroid volume by US and PET were made on 15 patients (two male and thirteen female, ages 28-73 years) with suspected Graves' disease, This comparison showed that in normal sized and enlarged thyroid glands (n=13), the ratio of functional PET to anatomical US volume was approximately 2:3, However, using the same ellipsoid model, PET and US assessment of anatomical volume agreed within the measurement errors. Owing to the presence of nodules and non-uniform distribution of radioiodine, the functional PET volume and anatomical US volume are often not equivalent. If high-resolution emission tomography (e.g. FET) is unavailable, the comparative data presented in this paper could be used to derive the functional volume from the US volume for calculating functional thyroid dose in hyperthyroid patients undergoing radioiodine therapy.
引用
收藏
页码:1470 / 1478
页数:9
相关论文
共 25 条
[1]  
[Anonymous], STAT METHODS MED RES
[2]  
Berg GEB, 1996, J NUCL MED, V37, P228
[3]  
BOCKISCH A, 1993, J NUCL MED, V34, P1632
[4]   VOLUMETRIC-ANALYSIS OF THYROID LOBES BY REAL-TIME ULTRASOUND [J].
BRUNN, J ;
BLOCK, U ;
RUF, G ;
BOS, I ;
KUNZE, WP ;
SCRIBA, PC .
DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 1981, 106 (41) :1338-1340
[5]   THYROID IMAGING USING POSITRON EMISSION TOMOGRAPHY - A COMPARISON WITH ULTRASOUND IMAGING AND CONVENTIONAL SCINTIGRAPHY IN THYROTOXICOSIS [J].
FLOWER, MA ;
IRVINE, AT ;
OTT, RJ ;
KABIR, F ;
MCCREADY, VR ;
HARMER, CL ;
SHARMA, HL ;
SMITH, AG .
BRITISH JOURNAL OF RADIOLOGY, 1990, 63 (749) :325-330
[6]  
FLOWER MA, 1994, EUR J NUCL MED, V21, P531
[7]   THE DETERMINATION OF THYROID VOLUME BY ULTRASOUND AND ITS RELATIONSHIP TO BODY-WEIGHT, AGE, AND SEX IN NORMAL SUBJECTS [J].
HEGEDUS, L ;
PERRILD, H ;
POULSEN, LR ;
ANDERSEN, JR ;
HOLM, B ;
SCHNOHR, P ;
JENSEN, G ;
HANSEN, JM .
JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 1983, 56 (02) :260-263
[8]  
Hertz Saul, 1942, JOUR CLIN INVEST, V21, P624
[9]   QUANTITATION IN POSITRON EMISSION COMPUTED-TOMOGRAPHY .1. EFFECT OF OBJECT SIZE [J].
HOFFMAN, EJ ;
HUANG, SC ;
PHELPS, ME .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1979, 3 (03) :299-308
[10]   LOW-DOSE RADIOIODINE GIVEN 6-MONTHLY IN GRAVES-DISEASE [J].
HOSKIN, PJ ;
MCCREADY, VR ;
HARMER, CL ;
SPATHIS, GS ;
COSGROVE, DO .
JOURNAL OF THE ROYAL SOCIETY OF MEDICINE, 1985, 78 (11) :893-898