Single trocar laparoscopic nephrectomy using magnetic anchoring and guidance system in the porcine model

被引:26
作者
Zeltser, Ilia S.
Bergs, Richard
Fernandez, Raul
Baker, Linda
Eberhart, Robert
Cadeddu, Jeffrey A.
机构
[1] SW Texas State Univ, Med Ctr, Dept Urol, Dallas, TX 75390 USA
[2] SW Texas State Univ, Clin Ctr Minimally Invas Urol Canc Treatment, Dept Urol, Dallas, TX USA
[3] SW Texas State Univ, Med Ctr, Dept Surg, Dallas, TX USA
[4] Univ Texas, Automat & Robot Res Inst, Arlington, TX 76019 USA
关键词
kidney; nephrectomy; laparoscopy; instrumentation; swine;
D O I
暂无
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Purpose: We assessed the feasibility of single keyhole laparoscopic surgery using a novel transabdominal magnetic anchoring and guidance system platform in the porcine model. Materials and Methods: A collaborative research group was formed to build a prototype system of magnetically anchored instruments for trocar-free lapa.roscopy. The design mandate was that the developed technology should be able to deploy into the insufflated abdomen through an existing 12 mm diameter trocar and then be moved into position in the peritoneum by manipulating external magnets. The magnetic anchoring and guidance system concept was advanced to a working prototype with a system of external magnetic anchors, an internal camera system and a hook cautery supported by an intra-abdominal robotic arm. This prototype system was then evaluated in vivo in a porcine laparoscopic nephrectomy model. Results: Two nonsurvival porcine laparoscopic nephrectomies were successfully completed without complications via a single 15 mm transumbilical trocar using the prototype magnetic anchoring and guidance system camera and the magnetically anchored robotic arm cauterizer. A conventional laparoscopic grasper was used for retraction through the 15 mm trocar after magnetic anchoring and guidance system deployment. The renal artery and vein were transected with a conventional Endo-GIA (R) stapler introduced through the 15 mm trocar. Procedure time was not recorded and blood loss was minimal. Conclusions: Single trocar laparoscopic nephrectomy using magnetically anchored instrumentation is technically feasible, demonstrating that intracorporeal. instrument manipulation may overcome the limitations of current laparoscopic and. robotic surgery by allowing unhindered intra-abdominal movement. This single access technique may be used with natural orifice surgery approaches and it has the potential to realize incision-free intra-abdominal surgery.
引用
收藏
页码:288 / 291
页数:4
相关论文
共 15 条
[1]   Progress in gastrointestinal tract surgery: the impact of gastrointestinal endoscopy - Marks lecture [J].
Beger, HG ;
Schwarz, A ;
Bergmann, U .
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2003, 17 (02) :342-350
[2]   Advantages and limits of robot-assisted laparoscopic surgery - Preliminary experience [J].
Corcione, F ;
Esposito, C ;
Cuccurullo, D ;
Settembre, A ;
Miranda, N ;
Amato, F ;
Pirozzi, F ;
Caiazzo, P .
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2005, 19 (01) :117-119
[3]   Cosmesis and body image after laparoscopic-assisted and open ileocolic resection for Crohn's disease [J].
Dunker, MS ;
Stiggelbout, AM ;
van Hogezand, RA ;
Ringers, J ;
Griffioen, G ;
Bemelman, WA .
SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES, 1998, 12 (11) :1334-1340
[4]   Transvaginal laparoscopic nephrectomy: Development and feasibility in the porcine model [J].
Gettman, MT ;
Lotan, Y ;
Napper, CA ;
Cadeddu, JA .
UROLOGY, 2002, 59 (03) :446-450
[5]   Robot-assisted abdominal surgery [J].
Gutt, CN ;
Oniu, T ;
Mehrabi, A ;
Kashfi, A ;
Schemmer, P ;
Büchler, MW .
BRITISH JOURNAL OF SURGERY, 2004, 91 (11) :1390-1397
[6]   Flexible transgastric peritoneoscopy: a novel approach to diagnostic and therapeutic interventions in the peritoneal cavity [J].
Kalloo, AN ;
Singh, VK ;
Jagannath, SB ;
Niiyama, H ;
Hill, SL ;
Vaughn, CA ;
Magee, CA ;
Kantsevoy, SV .
GASTROINTESTINAL ENDOSCOPY, 2004, 60 (01) :114-117
[7]   Transvesical endoscopic peritoneoscopy:: A novel 5 mm port for intra-abdominal scarless surgery [J].
Lima, Estevao ;
Rolanda, Carla ;
Pego, Jose M. ;
Henriques-Coelho, Tiago ;
Silva, David ;
Carvalho, Jose L. ;
Correia-Pinto, Jorge .
JOURNAL OF UROLOGY, 2006, 176 (02) :802-805
[8]   Symptomatic port-site hernia associated with a non-bladed trocar after laparoscopic live-donor nephrectomy [J].
Lowry, PS ;
Moon, TD ;
D'Alessandro, A ;
Nakada, SY .
JOURNAL OF ENDOUROLOGY, 2003, 17 (07) :493-494
[9]  
Marcovici I, 2001, JSLS, V5, P293
[10]  
Marohn Michael R, 2004, Curr Surg, V61, P466, DOI 10.1016/j.cursur.2004.03.009