Direct intra-aneurysm sac pressure measurement using tip-pressure sensors: In vivo and in vitro evaluation

被引:26
作者
Dias, NV
Ivancev, K
Malina, M
Hinnen, JW
Visser, M
Lindblad, B
Sonesson, B
机构
[1] Malmo Univ Hosp, Endovasc Ctr, Dept Radiol, Malmo, Sweden
[2] Malmo Univ Hosp, Endovasc Ctr, Dept Vasc Dis, Malmo, Sweden
[3] Leiden Univ, Med Ctr, Dept Surg, Vasc Surg Sect, Leiden, Netherlands
关键词
D O I
10.1016/j.jvs.2004.07.045
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objective: Direct intra-aneurysm sac pressure measurement with percutaneous translumbar puncture is a new method for follow-up after endovascular aneurysm repair. The purpose of this study was to evaluate a tip-pressure sensor system for intra-aneurysm pressure measurement in an in vitro aneurysm model and in vivo in patients by studying intraobserver variability. Methods. We used 0.014-inch guide wire-mounted tip-pressure sensors. For the in vitro aneurysm model, saccular aneurysms filled with thrombus were inserted in a left-heart-driven aneurysm model. Pressure was measured simultaneously with guide wire pressure sensors in the lumen of the model and within the aneurysm thrombus. In vivo, intraobserver variability was evaluated with double percutaneous translumbar puncture of the abdominal aortic aneurysm (AAA) with pressure measurement in 15 patients (14 men, 1 woman; median age, 75 years [63-80 years]; median AAA diameter, 55 mm [47-80 mm]) at a median of 32 months (2-100 months) after endovascular aneurysm repair. Mean pressure index was calculated as the percentage of mean intraaneurysm pressure relative to simultaneous mean systemic pressure. Results: In vitro, the difference in pressure between the tip-sensor measurements and the pressure output of the aneurysm model was 2 mm Hg (1-4 mm Hg) when the output varied between 150/50 and 200/100 mm Hg (n = 90). Mean pressure in the lumen of the model and within the aneurysm thrombus differed by I mm Hg (-5-15 turn Hg (n = 10). In vivo, intraobserver variability of mean pressure index (Bland-Altman plot) was 0% (-7%-17%; n = 15%). Conclusion: Direct intra-aneurysm sac pressure measurement with tip-pressure sensors mounted on 0.014-inch guide wires is a reliable and reproducible technique for measuring intra-AAA pressure both in vitro and in vivo.
引用
收藏
页码:711 / 716
页数:6
相关论文
共 12 条
[1]   Cellular content and permeability of intraluminal thrombus in abdominal aortic aneurysm [J].
Adolph, R ;
Vorp, DA ;
Steed, DL ;
Webster, MW ;
Kameneva, MV ;
Watkins, SC .
JOURNAL OF VASCULAR SURGERY, 1997, 25 (05) :916-926
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   Biomechanics of abdominal aortic aneurysm in the presence of endoluminal thrombus: Experimental characterisation and structural static computational analysis [J].
Di Martino, E ;
Mantero, S ;
Inzoli, F ;
Melissano, G ;
Astore, D ;
Chiesa, R ;
Fumero, R .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 1998, 15 (04) :290-299
[4]   Intra-aneurysm sac pressure measurements after endovascular aneurysm repair: Differences between shrinking, unchanged, and expanding aneurysms with and without endoleaks [J].
Dias, NV ;
Ivancev, K ;
Malina, M ;
Resch, T ;
Lindblad, B ;
Sonesson, B .
JOURNAL OF VASCULAR SURGERY, 2004, 39 (06) :1229-1235
[5]  
Kalvesten E, 1998, MICRO ELECTRO MECHANICAL SYSTEMS - IEEE ELEVENTH ANNUAL INTERNATIONAL WORKSHOP PROCEEDINGS, P574
[6]   Reduced pulsatile wall motion of abdominal aortic aneurysms after endovascular repair [J].
Malina, M ;
Länne, T ;
Ivancev, K ;
Lindblad, B ;
Brunkwall, J .
JOURNAL OF VASCULAR SURGERY, 1998, 27 (04) :624-631
[7]   Endoleakage after stent-graft treatment of abdominal aneurysm: Implications on pressure and imaging - an in vitro study [J].
Schurink, GWH ;
Aarts, NJM ;
Wilde, J ;
van Baalen, JM ;
Chuter, TAM ;
Kool, LJ ;
van Bockel, JH .
JOURNAL OF VASCULAR SURGERY, 1998, 28 (02) :234-241
[8]   Intra-aneurysm pressure measurements in successfully excluded abdominal aortic aneurysm after endovascular repair [J].
Sonesson, B ;
Dias, N ;
Malina, M ;
Olofsson, P ;
Griffin, D ;
Lindblad, B ;
Ivancev, K .
JOURNAL OF VASCULAR SURGERY, 2003, 37 (04) :733-738
[9]  
SZILAGYI DE, 1977, ARCH SURG-CHICAGO, V112, P399
[10]  
Vallabhaneni SR, 2003, J ENDOVASC THER, V10, P524, DOI 10.1583/1545-1550(2003)010<0524:CIPMRP>2.0.CO