HemoCue®, an accurate bedside method of hemoglobin measurement?

被引:47
作者
Rippmann, CE [1 ]
Nett, PC [1 ]
Popovic, D [1 ]
Seifert, B [1 ]
Pasch, T [1 ]
Spahn, DR [1 ]
机构
[1] Univ Zurich Hosp, Inst Anasthesiol, CH-8091 Zurich, Switzerland
关键词
hemoglobin measurement; HemoCue (R); hemoglobinometer; CO-Oximeter; spectrophotometer;
D O I
10.1023/A:1007451611748
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Objective. Evaluate the accuracy of this bedside method to determine hemoglobin (Hb) concentration in general surgery over a wide range of Hb values and to determine potential sources of error. Methods. Accuracy of Hb measurement using HemoCue(R) (AB Leo Diagnostics, Helsinborg, Sweden) was assessed in 140 surgical blood samples using 7 HemoCue(R) devices in comparison with a CO-Oximeter (IL 482, Instrumentation Laboratory, Lexington, MA). To analyze potential sources of error, packed red cells and fresh frozen plasma were reconstituted to randomized Hb levels of 2 - 18 g/dL. Results. In the surgical blood samples, the Hb concentration determined by the CO-Oximeter (HbCOOX) ranged from 5.1 to 16.7 g/dL and the Hb concentration measured by HemoCue(R) (HbHC) from 4.7 to 16.0 g/dL. Bias (HbCOOX - HbHC) between HbCOOX and HbHC was 0.6 +/- 0.6 g/dL (mean +/- SD) or 5.4 +/- 5.0 % (p < 0.001). Also in the reconstituted blood, the bias between HbCOOX and HbHC was significant (0.2 +/- 0.3 g/dL or 2.1 +/- 3.2 %; p < 0.001). The microcuvette explained 68 % of the variability between HbCOOX and HbHC. HemoCue thus underestimates the Hb concentration by 2 - 5 % and exhibits a 8 - 10 times higher variability with only 86.4 % of HbHC being within +/- 10 % of HbCOOX. Conclusion. Although the mean bias between HbCOOX and HbHC was relatively low, Hb measurement by HemoCue(R) exhibited a significant variability. Loading multiple microcuvettes and averaging the results may increase the accuracy of Hb measurement by HemoCue(R).
引用
收藏
页码:373 / 377
页数:5
相关论文
共 14 条