The weighted diagnostic distortion (WDD) measure for ECG signal compression

被引:295
作者
Zigel, Y
Cohen, A
Katz, A
机构
[1] Ben Gurion Univ Negev, Dept Elect & Comp Engn, IL-84105 Beer Sheva, Israel
[2] Ben Gurion Univ Negev, Fac Hlth Sci, Dept Cardiol, IL-84105 Beer Sheva, Israel
关键词
electrocardiogram; ECG compression; distortion measures; PRD; PQRST complex; weighted diagnostic distortion;
D O I
10.1109/TBME.2000.880093
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this paper, a new distortion measure for electrocardiogram (ECG) signal compression, called weighted diagnostic distortion (WDD) is introduced, The WDD measure is designed for comparing the distortion between original ECG signal and reconstructed ECG signal (after compression). The WDD is based on PQRST complex diagnostic features (such as P wave duration, QT interval, T shape, ST elevation) of the original ECG signal and the reconstructed one, Unlike other conventional distortion measures [e,g, percentage root mean square (rms) difference, or PRD], the WDD contains direct diagnostic information and thus is more meaningful and useful. Four compression algorithms were implemented (AZTEC, SAPA2, LTP, ASEC) in order to evaluate the WDD. A mean opinion score (MOS) test was applied to test the quality of the reconstructed signals and to compare the quality measure (MOSerror) with the proposed WDD measure and the popular PRD measure. The evaluators in the MOS test were three independent expert cardiologists, who studied the reconstructed ECG signals in a blind and a semiblind tests. The correlation between the proposed WDD measure and the MOS test measure (MOSerror) was found superior to the correlation between the popular PRD measure and the MOSerror.
引用
收藏
页码:1422 / 1430
页数:9
相关论文
共 41 条
[1]   A NEW DATA-REDUCTION ALGORITHM FOR REAL-TIME ECG ANALYSIS [J].
ABENSTEIN, JP ;
TOMPKINS, WJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1982, 29 (01) :43-48
[2]  
[Anonymous], 1986, BIOMEDICAL SIGNAL PR
[3]  
BARLAS GD, 1996, IEEE T BIOMED ENG, V43
[4]   Compression of multichannel ECG through multichannel long-term prediction [J].
Cohen, A ;
Zigel, Y .
IEEE ENGINEERING IN MEDICINE AND BIOLOGY MAGAZINE, 1998, 17 (01) :109-115
[5]  
Cohen A., 1990, P IEEE 90 S A S COMM, P45
[6]   AZTEC A PREPROCESSING PROGRAM FOR REAL-TIME ECG RHYTHM ANALYSIS [J].
COX, JR ;
NOLLE, FM ;
FOZZARD, HA ;
OLIVER, GC .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1968, BM15 (02) :128-&
[7]  
Draper N. R., 1966, APPL REGRESSION ANAL
[8]   AN ADAPTIVE REAL-TIME ECG COMPRESSION ALGORITHM WITH VARIABLE THRESHOLD [J].
FURHT, B ;
PEREZ, A .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (06) :489-494
[9]  
Gersho A., 1992, VECTOR QUANTIZATION
[10]   DETECTION OF THE P-WAVE AND T-WAVES IN AN ECG [J].
GRITZALI, F ;
FRANGAKIS, G ;
PAPAKONSTANTINOU, G .
COMPUTERS AND BIOMEDICAL RESEARCH, 1989, 22 (01) :83-91