Improvement of reproducibility of macular volume measurements using the Heidelberg retinal tomograph

被引:12
作者
Ang, A [1 ]
Tong, L [1 ]
Vernon, SA [1 ]
机构
[1] Univ Nottingham Hosp, Queens Med Ctr, Dept Ophthalmol, Nottingham NG7 2UH, England
关键词
D O I
10.1136/bjo.84.10.1194
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Aims-To develop a more reproducible method of macular volumetric analysis in order to facilitate serial monitoring of changes in retinal oedema with time. Methods-The Heidelberg retinal tomography (HRT) machine was used to scan the macula of 20 normal subjects and 40 diabetics with macular oedema. The volume above reference plane (VARP) within a 2 nun diameter and 3 mm diameter circle was measured twice in each eye. The position of the circle to be measured was carefully defined relative to major retinal vessels. As a modification to the previously published technique, fluctuation of the height of the contour Line relative to the focal plane of the eye is compensated for by an adjustment of the reference plane. The position of the circle was controlled relative to major retinal vessels using an acetate. The reproducibility of VARP measurement was assessed by three variables-the standard error of the difference on two isolated VARP measurements, the standard error of the difference between the average of three readings on two different occasions, as well, as the coefficient of variation (COV). Results-Both the 2 nun and 3 mm circles showed good reproducibility in VARP measurements. In normal subjects, the COV of VARP for 2 nun circles and 3 mm circles were 13% and 12% respectively. In patients with macular oedema, the COV of VARP for 2 nun circles and 3 mm circles were 9% and 11% respectively. Conclusion-The modification of VARP measurements between scans of the same eye has improved the COV from 31% to 9% in eyes with diabetic macular oedema.
引用
收藏
页码:1194 / 1197
页数:4
相关论文
共 7 条
[1]   QUANTITATIVE ASSESSMENT OF MACULAR EDEMA WITH OPTICAL COHERENCE TOMOGRAPHY [J].
HEE, MR ;
PULIAFITO, CA ;
WONG, C ;
DUKER, JS ;
REICHEL, E ;
RUTLEDGE, B ;
SCHUMAN, JS ;
SWANSON, EA ;
FUJIMOTO, JG .
ARCHIVES OF OPHTHALMOLOGY, 1995, 113 (08) :1019-1029
[2]   Scanning laser tomography Z profile signal width as an objective index of macular retinal thickening [J].
Hudson, C ;
Flanagan, JG ;
Turner, GS ;
McLeod, D .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1998, 82 (02) :121-130
[3]  
MENEZES AV, 1995, OPHTHALMOLOGY, V102, P230
[4]  
ZAMBARAKJI H, IN PRESS DOC OPHTHAL
[5]   Reproducibility of volumetric measurements of normal maculae with the Heidelberg retina tomograph [J].
Zambarakji, HJ ;
Evans, JE ;
Amoaku, WMK ;
Vernon, SA .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1998, 82 (08) :884-891
[6]   Assessment of the Heidelberg Retina Tomograph in the detection of sight-threatening diabetic maculopathy [J].
Zambarakji, HJ ;
Butler, TKH ;
Vernon, SA .
EYE, 1999, 13 (2) :136-144
[7]  
Zeimer R, 1996, INVEST OPHTH VIS SCI, V37, P1994