Performance of amperometric sensors based on multiple microelectrode arrays

被引:59
作者
Morf, WE [1 ]
de Rooij, NF [1 ]
机构
[1] Univ Neuchatel, Inst Microtechnol, CH-2000 Neuchatel, Switzerland
关键词
microelectrode array; steady-state current; dynamic response;
D O I
10.1016/S0925-4005(97)00159-7
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The steady-state current response and the dynamic characteristics (response to potential or concentration changes) of multiple microelectrode arrays are discussed. As a rule, an array with a low packing density of electrodes yields the ideal multiple response of a single microelectrode. In contrast. a closely packed array simply mimics the behavior of a macroelectrode of the same total surface area. The theoretical results are confirmed by experimental data. (C) 1997 Elsevier Science S.A.
引用
收藏
页码:538 / 541
页数:4