Comparison of three manual ventilation devices using an intubated mannequin

被引:59
作者
Hussey, SG [1 ]
Ryan, CA [1 ]
Murphy, BP [1 ]
机构
[1] Erinville Hosp, Dept Paediat & Child Hlth, Cork, Ireland
来源
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION | 2004年 / 89卷 / 06期
关键词
D O I
10.1136/adc.2003.047712
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Objective: To compare three devices for manual neonatal ventilation. Design: Participants performed a two minute period of ventilation using a self inflating device, an anaesthesia bag with attached manometer, and a Neopuff device. An intubated neonatal mannequin, approximating a 1 kg infant with functional lungs, was used for the study. Target ventilation variables included a rate of 40 breaths per minute, peak inspiratory pressure ( PIP) of 20 cm H2O, and positive end expiratory pressure (PEEP) of 4 cm H2O. The circuit was attached to a laptop computer for data recording. Results: Thirty five participants were enrolled, including consultant neonatologists, paediatricians, and anaesthetists, paediatric and anaesthetic registrars, and neonatal nurses. The maximum PIP recorded using the self inflating bag, anaesthetic bag, and Neopuff device were 75.9, 35.5, and 22.4 cm H2O respectively. There were significant differences between the devices for mean PIP (30.7, 18.1, and 20.1 cm H2O), mean PEEP (0.2, 2.8, and 4.4 cm H2O), mean airway pressure (7.6, 8.5, and 10.9 cm H2O), % total breaths less than or equal to 21 cm H2O PIP (39%, 92%, and 98%), and % total breaths greater than or equal to 30 cm H2O PIP (45%, 0, and 0). There was no difference between doctors and allied health professionals for the variables examined. Conclusion: The anaesthetic bag with manometer and Neopuff device both facilitate accurate and reproducible manual ventilation. Self inflating devices without modifications are not as consistent by comparison and should incorporate a manometer and a PEEP device, particularly when used for resuscitation of very low birthweight infants.
引用
收藏
页码:F490 / F493
页数:4
相关论文
共 27 条
[1]  
*AM AC PED AM HEAR, 2000, NEON RES PROGR TXB N, P3
[2]  
Brion L P, 1992, J Perinatol, V12, P281
[3]   Video recording as a means of evaluating neonatal resuscitation performance [J].
Carbine, DN ;
Finer, NN ;
Knodel, E ;
Rich, W .
PEDIATRICS, 2000, 106 (04) :654-658
[4]   Comparison of methods of bag and mask ventilation for neonatal resuscitation [J].
Finer, NN ;
Wade, RA ;
Craft, A ;
Henderson, C .
RESUSCITATION, 2001, 49 (03) :299-305
[5]  
FINER NN, 1986, PEDIATRICS, V77, P417
[6]   TREATMENT OF IDIOPATHIC RESPIRATORY-DISTRESS SYNDROME WITH CONTINUOUS POSITIVE AIRWAY PRESSURE [J].
GREGORY, GA ;
KITTERMAN, JA ;
PHIBBS, RH ;
TOOLEY, WH ;
HAMILTON, WK .
NEW ENGLAND JOURNAL OF MEDICINE, 1971, 284 (24) :1333-+
[7]  
HENLEY J, 2003, BRIT J RESUSCITATION, V2, P6
[8]   CHEST WALL RESTRICTION LIMITS HIGH AIRWAY PRESSURE-INDUCED LUNG INJURY IN YOUNG-RABBITS [J].
HERNANDEZ, LA ;
PEEVY, KJ ;
MOISE, AA ;
PARKER, JC .
JOURNAL OF APPLIED PHYSIOLOGY, 1989, 66 (05) :2364-2368
[9]   PERFORMANCE OF PEDIATRIC RESUSCITATION BAGS ASSESSED WITH AN INFANT LUNG SIMULATOR [J].
KAIN, ZN ;
BERDE, CB ;
BENJAMIN, PK ;
THOMPSON, JE .
ANESTHESIA AND ANALGESIA, 1993, 77 (02) :261-264
[10]   EVALUATION OF MASK-BAG VENTILATION IN RESUSCITATION OF INFANTS [J].
KANTER, RK .
AMERICAN JOURNAL OF DISEASES OF CHILDREN, 1987, 141 (07) :761-763