Newborn screening for hepatorenal tyrosinemia by tandem mass spectrometry: analysis of succinylacetone extracted from dried blood spots

被引:66
作者
Allard, P
Grenier, A
Korson, MS
Zytkovicz, TH
机构
[1] Tufts New England Med Ctr, Div Genet, Boston, MA 02111 USA
[2] Tufts Univ, Sch Med, Boston, MA 02111 USA
[3] Univ Laval, Ctr Med, Dept Med Genet, St Foy, PQ G1V 4G2, Canada
[4] Univ Massachusetts, Sch Med, New England Newborn Screening Program, Jamaica Plain, MA 02130 USA
关键词
tyrosinemia; succinylacetone; newborn screening; dried blood spots; filter paper; tandem mass spectrometry;
D O I
10.1016/j.clinbiochem.2004.07.006
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: To develop a method for the determination of succinylacetone (SA) in dried blood spots (DBS) using tandem mass spectrometry (MS/MS). Methods: SA was extracted from DBS with an acetonitrile and water solution (80:20 by volume) containing formic acid and hydrazine hydrate (both at 0.1% by Volume), and analyzed by MS/MS with a total run time per sample under 2 min. The reference range for SA in newborns was determined by analyzing a control group of 3199 DBS. SA was also measured in stored newborn specimens from three patients diagnosed clinically with hepatorenal tyrosinernia (FIT). Results: The within-run precision was less than or equal to7.4% and total precision was less than or equal to 12.2% on blood spots fortified with SA at 2, 10, and 50 [mumol/L. The limit of quantitation was 1 mumol/L and the calibration was linear from 1 to 50 mumol/L. A comparison of SA analysis of this MS/MS method with an established enzyme assay indirectly quantitating SA by inhibition of delta-aminolevulinic acid clehydratase demonstrated a strong con-elation. The reference range in the control group of unaffected newborn was determined to be <2 mumol/L, while SA in retrieved DBS Frorri FIT patients was significantly elevated, measuring 46.7, 36.5. and 23.2 mumol/L. Conclusions: We report a simple method that can be Used for screening HT in newborns. Based on our limited experience, a 2 mumol/L Cutoff could result in up to 100% sensitivity and specificity. (C) 2004 The Canadian Society of Clinical Chemists. All rights reserved.
引用
收藏
页码:1010 / 1015
页数:6
相关论文
共 21 条
[1]   Long-term therapy with NTBC and tyrosine-restricted diet in a murine model of hereditary tyrosinemia type I [J].
Al-Dhalimy, M ;
Overturf, K ;
Finegold, M ;
Grompe, M .
MOLECULAR GENETICS AND METABOLISM, 2002, 75 (01) :38-45
[2]   BIOCHEMICAL-STUDIES ON THE ENZYMATIC DEFICIENCIES IN HEREDITARY TYROSINEMIA [J].
BERGER, R ;
VANFAASSEN, H ;
SMITH, GPA .
CLINICA CHIMICA ACTA, 1983, 134 (1-2) :129-141
[3]   Sequential deglycosylation and utilization of the N-linked, complex-type glycans of human α1-acid glycoprotein mediates growth of Streptococcus oralis [J].
Byers, HL ;
Tarelli, E ;
Homer, KA ;
Beighton, D .
GLYCOBIOLOGY, 1999, 9 (05) :469-479
[4]   PITFALLS IN THE INITIAL DIAGNOSIS OF TYROSINEMIA - 3 CASE-REPORTS AND A REVIEW OF THE LITERATURE [J].
GOULDEN, KJ ;
MOSS, MA ;
COLE, DEC ;
TITHECOTT, GA ;
CROCKER, JFS .
CLINICAL BIOCHEMISTRY, 1987, 20 (03) :207-212
[5]   DETECTION OF SUCCINYLACETONE AND THE USE OF ITS MEASUREMENT IN MASS-SCREENING FOR HEREDITARY TYROSINEMIA [J].
GRENIER, A ;
LESCAULT, A ;
LABERGE, C ;
GAGNE, R ;
MAMER, O .
CLINICA CHIMICA ACTA, 1982, 123 (1-2) :93-99
[6]  
GRENIER A, 1985, METHOD ENZYMAT AN, V8, P73
[7]   Tyrosinaemia type I and NTBC (2-(2-nitro-4-trifluoromethylbenzoyl)-1,3-cyclohexanedione) [J].
Holme, E ;
Lindstedt, S .
JOURNAL OF INHERITED METABOLIC DISEASE, 1998, 21 (05) :507-517
[8]   STABLE ISOTOPE-DILUTION ANALYSIS OF SUCCINYLACETONE USING ELECTRON-CAPTURE NEGATIVE-ION MASS FRAGMENTOGRAPHY - AN ACCURATE APPROACH TO THE PRENATAL AND NEONATAL DIAGNOSIS OF HEREDITARY TYROSINEMIA TYPE-I [J].
JAKOBS, C ;
DORLAND, L ;
WIKKERINK, B ;
KOK, RM ;
DEJONG, APJM ;
WADMAN, SK .
CLINICA CHIMICA ACTA, 1988, 171 (2-3) :223-231
[9]   Liquid chromatography/tandem mass spectrometry methods for quantitation of mevalonic acid in human plasma and urine: method validation, demonstration of using a surrogate analyte, and demonstration of unacceptable matrix effect in spite of use of a stable isotope analog internal standard [J].
Jemal, M ;
Schuster, A ;
Whigan, DB .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2003, 17 (15) :1723-1734
[10]   The changing face of newborn screening: diagnosis of inborn errors of metabolism by tandem mass spectrometry [J].
Jones, PM ;
Bennett, MJ .
CLINICA CHIMICA ACTA, 2002, 324 (1-2) :121-128