Modelling of solid state nuclear track detector devices for radon measurements

被引:13
作者
Andriamanatena, R
Bacmeister, GU
Freyer, K
Ghose, R
Jonsson, G
Kleis, T
Treutler, HC
Enge, W
机构
[1] Univ Kiel, Inst Reine & Angew Kernphys, D-24118 Kiel, Germany
[2] Univ Autonoma Barcelona, Dept Fis, Bellaterra 089193, Spain
[3] UFZ Helmholtz Ctr Environm Res, Sekt Analyt, D-04318 Leipzig, Germany
[4] Lund Inst Technol, S-22100 Lund, Sweden
关键词
LR-115; track etching rate; bulk etching rate; radon measurements; EU-project;
D O I
10.1016/S1350-4487(97)00159-5
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Theoretical analytical calculations of the radon registration have been performed for the solid state nuclear track detectors (SSNTD) LR-115 Type II (peliculable) produced by Kodak-Pathe (France). Known and measured properties of LR-115 together with the well established etching theory using "Huygens' principle" were used to carry out these detailed calculations for different types of exposure devices for indoor room and outdoor soil measurements. The results of these modeling calculations are compared with experimental calibration measurements and show good agreement.
引用
收藏
页码:657 / 662
页数:6
相关论文
共 8 条
[1]   PRACTICAL AND THEORETICAL CALIBRATIONS OF RN-SENSITIVITIES FOR PLASTIC NUCLEAR TRACK DETECTORS [J].
ANDRIAMANANTENA, R ;
ENGE, W .
RADIATION MEASUREMENTS, 1995, 25 (1-4) :625-626
[2]  
ANDRIANAMANTENA R, 1995, THESIS U KIEL
[3]  
BARRILLON R, 1994, THESIS U BESANCON FR
[4]  
ENGE W, 1980, NUCL TRACKS RAD MEAS, V4, P283
[5]   RADON GAS - WHERE FROM AND WHAT TO DO [J].
JONSSON, G .
RADIATION MEASUREMENTS, 1995, 25 (1-4) :537-546
[6]  
KLEIS T, 1992, NUCL TRACKS RAD MEAS, V20, P495
[7]   COMPUTATION OF THE CALIBRATION FACTOR FOR THE CUP TYPE SSNTD RADON MONITOR [J].
SIMA, O .
RADIATION MEASUREMENTS, 1995, 25 (1-4) :603-606
[8]  
SOMOGYI G, 1986, E25 HUNG AC SCI I NU