Capillary blood sampling: the pain of single-use lancing devices

被引:7
作者
Fruhstorfer, H [1 ]
机构
[1] Univ Marburg, Inst Physiol, D-33033 Marburg, Germany
来源
EUROPEAN JOURNAL OF PAIN-LONDON | 2000年 / 4卷 / 03期
关键词
capillary blood sampling; lancing pain; mechanical trauma; skin; single-use device;
D O I
10.1053/eujp.2000.0179
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Six disposable automatic lancing devices were examined in 50 healthy adult volunteers with regard to pain intensity, blood volume and user safety. Devices with a trigger-linked lancer movement were significantly more painful than those with a preset lancet speed. Residual pain was more frequent after the use of devices with a blade compared to devices with a needle. Only half of the devices reliably produced sufficient blood Volumes (>50 mu l). Two of the devices could accidentally be re-used and constitute a hazard for patient and personnel. It is concluded that some of the disposable automatic lancing devices are safe and reliable but none of them causes minimal pain. (C) 2000 European Federation of Chapters of the international Association for the Study of Pain.
引用
收藏
页码:301 / 305
页数:5
相关论文
共 10 条
[1]  
BERDE C, 1993, IASP NEWSLETTER MAR, P2
[2]  
BLUMENFELD TA, 1979, LANCET, V1, P230
[3]   Capillary blood sampling: relation between lancet diameter, lancing pain and blood volume [J].
Fruhstorfer, H ;
Schmelzeisen-Redeker, G ;
Weiss, T .
EUROPEAN JOURNAL OF PAIN, 1999, 3 (03) :283-286
[4]  
Fruhstorfer H, 1995, PRACTICAL DIABETES, V12, P253
[5]  
GARVEY K, 1998, 9805 MDA
[6]   NEONATAL OSTEOMYELITIS OF CALCANEUS - COMPLICATION OF HEEL PUNCTURE [J].
LILIEN, LD ;
HARRIS, VJ ;
RAMAMURTHY, RS ;
PILDES, RS .
JOURNAL OF PEDIATRICS, 1976, 88 (03) :478-480
[7]   ALLEVIATION OF THE PAIN OF HEEL PRICK IN PRETERM INFANTS [J].
MCINTOSH, N ;
VANVEEN, L ;
BRAMEYER, H .
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 1994, 70 (03) :F177-F181
[8]  
MEITES S, 1988, CLIN CHEM, V34, P1890
[9]   NOSOCOMIAL TRANSMISSION OF HEPATITIS-B VIRUS ASSOCIATED WITH THE USE OF A SPRING-LOADED FINGERSTICK DEVICE [J].
POLISH, LB ;
SHAPIRO, CN ;
BAUER, F ;
KLOTZ, P ;
GINIER, P ;
ROBERTO, RR ;
MARGOLIS, HS ;
ALTER, MJ .
NEW ENGLAND JOURNAL OF MEDICINE, 1992, 326 (11) :721-725
[10]   FINGER SEPSIS - POSSIBLE COMPLICATION OF SELF MONITORING OF BLOOD-GLUCOSE CONCENTRATIONS [J].
RYAN, EA ;
MILLER, J ;
SKYLER, JS .
BRITISH MEDICAL JOURNAL, 1983, 286 (6378) :1614-1615