Practicality and accuracy of prehospital rapid venous blood glucose determination

被引:11
作者
Holstein, A
Kühne, D
Elsing, HG
Thiessen, E
Plaschke, A
Widjaja, A
Vogel, VY
Egberts, EH
机构
[1] Klinikum Lippe Detmold, Dept Med 1, D-32756 Detmoid, Germany
[2] Klinikum Lippe Detmold, Inst Anesthesiol, D-32756 Detmoid, Germany
[3] Orthoclin Diagnost GmbH, Neckargemund, Germany
[4] Med Hsch Hannover, Dept Endocrinol, Hannover, Germany
关键词
emergency medicine; diabetes; rapid blood glucose test; hypoglycemia; reflectometer;
D O I
10.1053/ajem.2000.7324
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Blood glucose testing plays an important role in emergency medicine. Although the use of visual reagent test strips is widely established in this setting, the accuracy of reflectometric blood glucose determinations under emergency conditions has rarely been investigated, In a prospective study, 522 of a total of 3,217 patients undergoing emergency blood glucose testing had parallel blood glucose measurements performed using a specific enzymatic method. These 522 patients (aged 61.4 years, 54% men, 90 cases of severe hypoglycemia) had an intravenous access placed at the scene of the emergency. Venous whole blood from the introducer needle of the access was applied to the test strip and the glucose measured with a GlucoTouch reflectometer (LifeScan, Inc.). A blood sample from the intravenous access was then immediately collected in a monovette for subsequent glucose determination in a chemical laboratory (hexokinase method) within 20 to 40 minutes. The emergency glucose measurements(mean: 7.3 mmol/L [95% confidence interval [CI] 6.9 to 7.7]; range: 0.55 to 27.7) correlated well with the reference laboratory results (Pearson's r =.98; linear regression analysis: slope 1.0, axial intercept 1.74). Error grid analysis also showed good agreement between corresponding measurements: zone A 96.7%, B 2.5%, C 0% and D 0.8%. The mean difference using the Bland-Altman method was 0.14 mmol/L; 2 SD 1.8 mmol/L; minimum -7.0 mmol/L; maximum 4.4 mmol/L. The accuracy of the rapid venous blood glucose determination by constantly changing emergency teams was high. Especially in 90 hypoglycemic patients, there were no deviations from the reference method that could have led to clinically relevant wrong decisions. The method of collecting whole blood directly from the venous access is simple and robust, and is independent of the hemodynamic status of the patient. (Am J Emerg Med 2000,18:690-694. Copyright (C) 2000 by W.B. Saunders Company).
引用
收藏
页码:690 / 694
页数:5
相关论文
共 17 条
  • [1] PENTAMIDINE-INDUCED DERANGEMENTS OF GLUCOSE-HOMEOSTASIS - DETERMINANT ROLES OF RENAL-FAILURE AND DRUG ACCUMULATION - A STUDY OF 128 PATIENTS
    ASSAN, R
    MAYAUD, C
    PERRONNE, C
    MATHERON, S
    ASSAN, D
    ZUCMAN, D
    CHOTARD, L
    [J]. DIABETES CARE, 1995, 18 (01) : 47 - 55
  • [2] Bachmann W, 1995, DIAB STOFFW, V4, P83
  • [3] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [4] A CLINICAL COMPARISON OF THE PERFORMANCE OF 4 BLOOD-GLUCOSE REAGENT STRIPS
    CHEELEY, RD
    JOYCE, SM
    [J]. AMERICAN JOURNAL OF EMERGENCY MEDICINE, 1990, 8 (01) : 11 - 15
  • [5] EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE
    CLARKE, WL
    COX, D
    GONDERFREDERICK, LA
    CARTER, W
    POHL, SL
    [J]. DIABETES CARE, 1987, 10 (05) : 622 - 628
  • [6] EBERT C, 1999, DIAB STOFFW, V8, P237
  • [7] HENRICHS H, 1997, DIABETES J, V5, P15
  • [8] Holstein A., 1996, Intensivmedizin und Notfallmedizin, V33, P594
  • [9] Jones Jonathan L., 1992, Journal of Emergency Medicine, V10, P679, DOI 10.1016/0736-4679(92)90524-W
  • [10] KOSCHINSKY T, 1998, DIABETESPROFI, V2, P10